MEDICAL ADVANCED TECHNOLOGY 1253 CORDLESS DRILL SYSTEM User’s Manual No. 6.
Intended use: The Drill 4-807 is used in bone surgery for drilling and tapping, inserting and removing screws and placing Kirschner wires. The drill has a cannulated drive shaft, so that drill wires can be briefly inserted and fed through the drill (diameter of cannulation: 3.2mm) for safe working. The tool holders (attachments) permit quick and easy chucking of the required tools. A choice of chucks and holders (attachments) ensure compatibility with various tool ranges.
Cleaning, disinfection, sterilisation: Cleaning and disinfection Wipe the surfaces immediately after use with a lint-free cloth moistened with a cleaning agent or disinfectant, or use a disinfectant spray, to prevent exudates or other contamination from drying on. For this manual wipe--down, use a cleaning agent or disinfectant approved for surface disinfection under national regulations (DGHM – German Association of Hygiene and Microbiology).
Transport and storage conditions: To prevent damage! The Cordless Drill 4-807 and attachments must be transported in such a way that no damage can occur. This particularly means ensuring suitable packaging. The same attention must be paid to correct storage. This device must only be kept in a dry, closed place! Safety instruction: Remove the battery from the Cordless Drill 1. Before sterilisation 2. If any fault occurs 3. For transport or storage when not in use 4.
Technical data: Type of equipment: Type BF Cordless Drill 4-807 Mono Block Battery oscillating Saw 4-808 Voltage 9,6 V 9,6 V Rated capacity of the Rechargeable Battery 4-810 1.2 Ah 1.2 Ah Speed, infinitely adjustable 0-750 rpm 0-14000 rpm Max. torque 4 Nm Weight incl. battery 1250 grams 1500 grams Motor output 70 W 70 W IP66 IP66 Type according to Directive: EN 60601-1-2 EN 60601-1 / IEC 601-1 MED.
Disposal: MAT GmbH & Co. KG will dispose of the Cordless Drill 4-807 and atttachments (without any environmental contamination)in accordance with the provisions of MED. UMW MAT GmbH & co.
OVERVIEW: 1 6. Clockwise rotation 7. Main switch and reverse rotation R L 2 1. Cordless Drill 2. Battery Compatment 3. Rechargeable Battery 4. End cap 5. Chuck with Drill Bit 6. Clockwise Rotation 7. Main switch and reverse Rotation 3 4 Important To obtain full power from the drill during the surgical operation, always use freshly charged batteries. Preparation Step 1. Open the end cap (4) by turning in the direction indicated by the arrow Step 2.
Overview of attachments: NOTE: All the clutches and chucks for the Cordless Drill system are mounted in the same way! • Step 1.
ATTACHMENTS FOR CORDLESS DRILL 4-807 ( GB ) Overview of attachments: Order number Description: 1.) REF. 4 – 809 Long rechargeable battery with extension for Cordless Drill 4-807 for increased power output 2.) REF. 4 – 810 Standard rechargeable battery 3.) REF. 4 – 811 Funnel for inserting Rechargeable Battery 4-809 or 4-810 into Cordless Drill 4-807 (to avoid contamination of the Cordless Drill. If it is not used, there may be a loss of sterility) 4.) REF.
Cleaning, disinfection and sterilisation of attachments: (except Rechargeable Battery 4-809/4-810 and Desktop Charger 4-814/4-816, because these products are not categorised as medical devices ) CLEANING AND DISINFECTION: CARE AND MAINTENANCE Care and maintenance are vital to keep the attachments in good condition and fully functional! Causes of corrosion with stainless steel used for instruments. Corrosion is damage or wear that can be caused by a chemical reaction, e.g. due to: 1.) Damaged surfaces 2.
1. MECHANICAL CLEANING: Mechanical cleaning is the preferred method. Strictly follow the instructions supplied by the manufacturer of the cleaning machine. Prepare the instruments and attachments for mechanical cleaning. Multipart instruments and attachments that can be dismantled must be taken apart or opened for cleaning. All tips and delicate parts should be cleaned manually. Do not overfill wire trays; place heavy parts on the bottom of the tray.
Instructions for use of attachments: Desktop chargers 4-814/4-816 Version: 28.04.2008 1. Main switch/pilot lamp <=> Switch is alight when the charger is ON 2. Charge indicator lamp <=> Glows steady red when a depleted RECHARGEABLE BATTERY 4-809/4-810 is being charged 2. Charge indicator lamp <=> Glows green when RECHARGEABLE BATTERY 4-809/4-810 is fully charged 3. Charging compartment <=> For inserting the RECHARGEABLE BATTERY 4-809/4-810 in the direction indicated by the arrow 4.
Care and maintenance: Unplug from the power supply before cleaning. Use a damp cloth for cleaning the Charger 4-814/4-816 and BATTERY 4-809/4-810. Do not use cleaning agents. Make sure that the contacts of BATTERY 4-809/4-810 are kept free of moisture. Similarly, protect the charging compartment (3) of the Charger 4-814/4-816 against moisture.Never immerse in liquid or clean under flowing water! Repair: MAT GmbH & Co.
Technical Data: Type of equipment: Chargers 4-814 and 4-816 Rated voltage 240 V and 115 V Rated frepuency 50 HZ Rated umput 14 VA Rated output voltage 12 V Rated output current 1100 mA max. Protection class II Degree of protection, as per DIN EN 529 IPX0 Function: The Chargers 4-814 and 4-816 are controlled and monitored by a microcontroller. Irrespective of the charge level of the battery, the controller measures the momentary capacity, and controls the charging process.
Notice: 15
MEDICAL ADVANCED TECHNOLOGY Cordless Drill System User’s Manual No. 6 ( GB ) © MAT MAT GmbH & Co.KG Friedrich-Wöhler Straße 10 78576 Emmingen-Liptingen, Germany Phone: +49 (0)7465 909 188 · Fax : +49 (0)7465 909 024 www.mat-medical.com · info@mat-medical.com Revision 06.04.