User's Manual
Table Of Contents
- 1Intended use
- 2Safety information
- 3Overview
- 3.1Introduction
- 3.2Product composition
- 3.2.1Main Box
- 3.2.2Nodes
- 3.2.3Straps
- 3.2.4Computer and Camera
- 3.3Working conditions
- 3.4Transport and storage conditions
- 3.5Symbols used in this product
- 4Specifications and parameters
- 4.1Specification
- 4.2 Indicator light description
- 4.2.1Main Box indicator
- 4.2.2Node indicator
- 5Instructions for use
- 5.1 Preparation for use
- 5.1.1Charging
- 5.1.2Device connection
- 5.1.3Venue set-up
- 5.2 Strap-on wear
- 5.2.1Effect after wearing
- 5.2.2Chest strap
- 5.2.3Waist strap
- 5.2.4Wrist strap
- 5.2.5Thigh strap
- 5.2.6Ankle strap
- 5.2.7Vamp strap
- 5.3Node Wearing
- 5.3.1Node snap-in
- 5.3.2Node pull-out
- 5.4Getting started with the software
- 5.4.1Visit the Wearable Motion and Gait Quantification
- 5.4.2New assessment
- 5.4.3Enter the assessment name and description of the a
- 5.4.4Add a check
- 5.4.5Select an assessment template to carry out the ass
- 5.4.6Enter patient information
- 5.4.7Start assessment
- 5.4.8View the assessment report
- 5.4.9Other functions
- 5.5After use
- 5.5.1End of use
- 6Cleaning and disinfection
- 6.1 Cleaning the main chassis and nodes
- 6.2 Cleaning straps
- 7Maintenance and Care
- 8Exceptions and their handling
- 9Cautions
- 10Contraindications
- 11Electromagnetic compatibility information
- 12Name and content of hazardous substances
- 13REGULATORY INFORMATION
- 14Quality commitment
- 15Disposal
- 16Copyright and Responsibility
- 24 -
humidity and exposure to the sun.
8 Exceptions and their handling
Type of fault
Possible causes and treatment
The main case charging indicator
does not light up
1 Verify that the power cable is connected properly
2 If the power cable is connected properly
Please call our customer service hotline for solutions.
The main chassis operating status
indicator does not light up
1 The data cable is not connected.
Connect the data cable correctly.
2 Data cable problem.
Verify that the data connection cable between the main case and the computer is intact.
3 If you confirm that the data cable is connected properly.
Please call our customer service hotline for solutions.
Node indicator does not light up
1 No power to the node.
Please charge the nodes.
2 If the node charge indicator also does not light up.
Please call our customer service hotline for solutions.
Node indicator flashes red
1 Node hardware exception.
Please charge the node, if it still flashes red, please call the customer
service hotline for a solution.
Unable to charge
1 Not properly charged
Check that the charging cable or charging base is correctly connected.
2 Battery or charging component failure
Please contact the manufacturer for a solution.
Table 10 Exceptions and their handling
9 Cautions
Please ensure that the product is fully charged before use. When the battery is low, stop using the
device for a short time and recharge the system as soon as possible.
Please follow the operating instructions when using the product.
Do not wash the straps in the washing machine, hand wash only.
10 Contraindications
In the case of patients with abnormal gait, please assess whether they are at risk of execution before
using the system, if so, do not carry out an assessment.