User's Manual

42
RIOTRONIC X+
User Manual
When charging is finished, user will be informed. Charger may be unplugged. Logger will be
automatically turned off.
Figure 38. Charging done message
6.8. Firmware uploading
To change the firmware, file FIRMWARE.BIN have to be placed inside BOOT folder of the
internal.
Logger’s memory. FIRMWARE.BIN file have to be encoded using device’s serial number.
To copy this file to Logger regular PC connection (if old firmware is working) or bootloader
mode can be used.
To start bootloader mode Logger must be turned off first. Then up and down arrow keys
have to be pressed and hold while ON/OFF is pressed to turn the device on. Logger starts in
bootloader mode, shows bootloader version, device’s serial number and allow to connect USB
cable to copy the file. After copying the file Logger have to be unmounted from the PC
system and turned off. When turned on again Logger looks in BOOT folder for new firmware.
If the file is OK and encoded with correct serial number, Logger will change content of the
firmware flash memory and the file will be renamed to FLASH.BIN.
6.9. Service menu
To get access to Service menu, user PIN must be set first. After Logger is turned on, service
PIN 62926082 must be entered instead of user PIN. Service PIN is fixed and cannot be
changed. There is a second stage of authorisation token checking. There are four digits
displayed on the screen. Another four digits must be entered. Token digits must be
calculated according to the algorithm below:
First two digits have to be multiplied and then third and fourth digit have to be
added. The result gives first two digits of token.
Digits three and four have to be multiplied and then first and second digit have to be
added.
The result gives third and fourth digit of token.