Operating instructions

Commissioning
50
Integration function
With this function, the computed value from the linearization table or the current flow value for
channel calculation, or that of the analog input can be numerically integrated to create a totalizer
for example.
The totalizer is calculated as follows:
The measuring interval is 0.1 s.
In most instances, the integration basis is the same time unit as the time basis of the signal to be
integrated.
Example: analog input l/s integration basis s
Simple preset counter
G09-RIA452ZZ-15-00-xx-en-010
Fig. 21: Simple preset counter
If the preset counter is activated, limit values 1 and 2 are permanently assigned to the preset
counter function (output 1 = main switchoff, output 2 = preliminary switchoff). Digital input 1
is permanently assigned to the "Reset and restart preset counter" function.
NFX V-weir
30
45
60
90
NFX triangular weir
30 = NFX triangular weir 30
45 = NFX triangular weir 45
60 = NFX triangular weir 60
90 = NFX triangular weir 90
Function
(menu
item)
Parameter
setting
Description