User guide
71M6533-DB Demo Board User’s Manual   
Page: 40 of 75  `  REV 3 
2.2.2  CALIBRATION PROCEDURE WITH FIVE MEASUREMENTS  
Each phase is calibrated individually. The calibration procedure is as follows: 
1)  The calibration factors for all phases are reset to their default values, i.e. CAL_In = CAL_Vn = 16384, 
and PHADJ_n = 0. 
2)  An RMS voltage V
ideal
 consistent with the meter’s nominal voltage is applied, and the RMS reading 
V
actual
 of the meter is recorded. The voltage reading error Axv is determined as 
Axv = (V
actual - 
V
ideal 
) / V
ideal
3)  Apply the nominal load current at phase angles 0°, 60°, 180° and –60° (-300°). Measure the Wh 
energy each time and record the errors E
0
, E
60
, E
180
, and E
300
.  
4)  Calculate the new calibration factors CAL_In, CAL_Vn, and PHADJ_n, using the formulae presented 
in section 2.1.2 or using the spreadsheet presented in section 2.2.4. 
5)  Apply the new calibration factors CAL_In, CAL_Vn, and PHADJ_n to the meter. The memory 
locations for these factors are given in section 1.9.1.  
6)  Test the meter at nominal current and, if desired, at lower and higher currents and various phase 
angles to confirm the desired accuracy. 
7)  Store the new calibration factors CAL_In, CAL_Vn, and PHADJ_n in the EEPROM memory of the 
meter. If a Demo Board is calibrated, the methods involving the command line interface shown in 
sections 1.9.3 and 1.9.4 can be used. 
8)  Repeat the steps 1 through 7 for each phase. 
9)  For added temperature compensation, read the value TEMP_RAW (CE RAM) and write it to 
TEMP_NOM (CE RAM). If Demo Code 4.6n or later is used, this will automatically calculate the 
correction coefficients PPMC and PPMC2 from the nominal temperature and from the characterization 
data contained in the on-chip fuses. 
Tip: Step 2 and the energy measurement at 0° of step 3 can be combined into one step. 
2.2.3  CALIBRATION PROCEDURE FOR ROGOWSKI COIL SENSORS 
Demo Code containing CE code that is compatible with Rogowski coils is available from MAXIM INTEGRADED 
PRODUCTS.  
Rogowski coils  generate a  signal that is the  derivative  of the current. The CE code  implemented in the 
Rogowski  CE image digitally  compensates  for  this effect  and  has  the usual gain  and  phase calibration 
adjustments. Additionally, calibration adjustments  are  provided  to eliminate voltage coupling  from  the sensor 
input. 
Current sensors  built from Rogowski coils have  a relatively  high output impedance that  is  susceptible to 
capacitive coupling from the large  voltages present in the meter.  The most dominant  coupling  is usually 
capacitance between the primary of the coil and the coil’s output. This coupling adds a component proportional 
to the derivative of voltage to the sensor output. This effect is compensated by the voltage coupling calibration 
coefficients. 
As with the CT procedure, the calibration procedure for Rogowski sensors uses the meter’s display to calibrate 
the voltage path and the pulse outputs to perform the remaining energy calibrations. The calibration procedure 
must be done to each phase separately, making sure that the pulse generator is driven by the accumulated real 
energy  for just  that  phase. In  other words,  the  pulse generator  input  should  be  set  to  WhA, WhB, or WhC, 
depending on the phase being calibrated. 
In preparation of the calibration, all calibration parameters are set to their default values. VMAX and IMAX are 
set to reflect the system design parameters. WRATE and PULSE_SLOW, PULSE_FAST are adjusted to obtain the 
desired Kh. 










