User guide
71M6533-DB Demo Board User’s Manual   
Page: 50 of 75  `  REV 3 
Figure 2-13: Meter with Calibration System 
Maxim’s Teridian Demo Boards are not calibrated prior to shipping. However, the Demo Board pulse outputs 
are tested and compared to the expected pulse output rate. Figure 2-14 shows the screen on the controlling PC 
for a typical Demo Board. The error numbers are given in percent. This means that for the measured Demo 
Board, the sum of all errors resulting from tolerances of PCB components, CTs, and 71M6533 tolerances was –
3.41%, a range that can easily be compensated by calibration. 
Figure 2-15 shows a load-line obtained with a 6533 in differential mode. As can be seen, dynamic ranges of 
10,000:1  for  current  can  be  achieved with  good  circuit  design,  layout,  cabling,  and,  of  course,  good current 
sensors. 
Figure 2-14: Calibration System Screen 
Calibrator
AC Voltage
Current CT
Meter 
under 
Test
Optical Pickup 
for Pulses
Calibrated 
Outputs
Pulse 
Counter
PC