User guide
71M6533-DB Demo Board User’s Manual   
Page: 51 of 75  `  REV 3 
Figure 2-15: Wh Load Line in Differential Mode at Room Temperature 
2.5.2  EEPROM 
Testing the EEPROM provided on  the Demo Board is straightforward and can be  done using the  serial 
command line interface (CLI) of the Demo Code. 
To write a string of text characters to the EEPROM and read it back, we apply the following sequence of CLI 
commands: 
>EEC1  Enables the EEPROM 
>EESthis is a test  Writes text to the buffer 
>EET80  Writes buffer to address 80 
Written to EEPROM address 00000080 74 68 69 73 20 69 73 20 61 ….   
Response from Demo Code 
>EER80.E  Reads text from the buffer 
Read from EEPROM address 00000080 74 68 69 73 20 69 73 20 61 ….   
Response from Demo Code 
>EEC0  Disables the EEPROM 
2.5.3  RTC 
Testing  the  RTC inside the  71M6533  IC  is  straightforward and  can  be  done  using  the  serial  command line 
interface (CLI) of the Demo Code. 
To set the RTC and check the time and date, we apply the following sequence of CLI commands: 
>M10  LCD display to show calendar date 
>RTD05.09.27.3  Sets the date to 9/27/2005 (Tuesday) 
>M9  LCD display to show time of day 
>RTT10.45.00  Sets the time to 10:45:00. AM/PM distinction: 1:22:33PM = 13:22:33 
-0.2 
-0.15 
-0.1 
-0.05 
0 
0.05 
0.1 
0.15 
0.2 
0.01  0.1  1  10  100  1000 
Error [%] 
I [A] 
Load Line in Differential Mode 
Error(%) 