Datasheet

Evaluates: MAX11044–MAX11049
MAX11046 Evaluation Kit
_______________________________________________________________________________________ 5
Data Logging
Start Convert
The Number of Samples drop-down list selects the
desired number of conversions. Press the Start
Convert button to start sampling at the rate set by the
Sample Rate edit box.
The on-board-generated convert start signal range is
from 1ksps to 130ksps. The external convert start signal
range is from 1ksps to 180ksps.
If using an external convert start signal, the GUI soft-
ware will measure and display the sample rate.
Output File Data
The Save to File button is not active until the sampling
is done. Press the Save to File button to save the sam-
pled data into a file.
Detailed Description of Hardware
MAX11046 EV Kit
The MAX11046 EV kit board provides a proven layout
for evaluating the MAX11046 8-channel, 16-bit, simulta-
neous-sampling ADC. The EV kit comes with the
MAX11046ETN+ installed. Eight BNC connectors are
provided on the EV kit for all input channels of the
MAX11046.
Power Supply
One DC power supply is required to power up the
MAX11046 EV kit. Connect the power supply (from 6V to
8V) between the VDD and the GND pads on the EV kit.
Jumper Settings
Table 1 provides an overview of jumper settings.
JUMPER
SHUNT
POSITION
DESCRIPTION
1-2*
BNC1 protected by a transient
voltage suppressor (D2)
2-3 BNC1 shorted to ground
JU1
Open BNC1 disconnected from D2
1-2*
BNC2 protected by a transient
voltage suppressor (D1)
2-3 BNC2 shorted to ground
JU2
Open BNC2 disconnected from D1
1-2* BNC1 bypassing optional op amp
JU3
2-3
BNC1 connected to an optional
op amp
1-2* BNC2 bypassing optional op amp
JU4
2-3
BNC2 connected to an optional
op amp
1-2
BNC1 connected to CH0 through a
10kΩ resistor
3-4
BNC1 connected to CH0 through a
1kΩ resistor
5-6
BNC1 connected to CH0 through a
100Ω resistor
JU5
7-8* BNC1 connected to CH0
JUMPER
SHUNT
POSITION
DESCRIPTION
1-2
BNC2 connected to CH1 through a
10kΩ resistor
3-4
BNC2 connected to CH1 through a
1kΩ resistor
5-6
BNC2 connected to CH1 through a
100Ω resistor
JU6
7-8* BNC2 connected to CH1
1-2*
Use on-board external voltage
reference (U3)
JU7
Open
U se an i nter nal vol tag e r efer ence or
user - sup p l i ed exter nal vol tag e
r efer ence connected to p i n 2 of JU 7
1-2*
Use on-board DSP (U2) generated
convert start signal
JU8
Open
Use a user-supplied external convert
start signal connected to pin 2 of
JU8
1-2* CH2 connected to BNC3 directly
2-3 CH2 shorted to ground
JU9
Open
CH2 connected to BNC3 through a
1kΩ resistor
Table 1. Jumper Configuration
*
Default position.
Note: CH_ refer to MAX11046 and MAX11049 pin names only.