Datasheet
Maxim Integrated
│
4
MAX31865 Evaluation Kit Evaluates: MAX31865
www.maximintegrated.com
Figure 1. MAX31865 EV Kit GUI (Registers Tab)
Detailed Description of Software
Device Conguration
Upon successful completion of the setup procedure, the
user is ready to begin evaluation of the device. The device
powers up in “normally off” conversion mode. To begin
conversions, first enable the VBIAS output by clicking
the On radio button located in the MAX31865 Setup Info
group box in the Registers
tab(Figure1).Next,enable
conversions by changing the Conversion Mode to Auto
by clicking the Auto radio button. It is now time to read the
results. This is accomplished by clicking the Read button
in the RTD Resistance Register group box. The conver-
sion results are displayed below the button in hex and
binary format. Also note that the ADC result is displayed
in decimal format. The RTD resistance is then calculated
from the ADC result and the value entered in the RREF
Nominal Resistance edit box located in the MAX31865
Setup Info
groupbox.Withthesupplied1kΩresistor,the
calculatedRTDresistanceshouldbe1kΩ±6%duetothe
tolerance of the provided resistors. Note: Since there is
not an RTD connected at this time, the RTD temperature
equations are not valid registers.