Oxygen Equipment User Manual

14
4. SENSOR REMOVAL AND REPLACEMENT
The OM-25ME is shipped with a new MAX-250E oxygen sensor installed.
Although the sensor has a very long expected life, eventually the sensor
will require replacement. Removing or installing a sensor, when
necessary, is a very simple procedure. To remove and install a new
sensor:
1) Grasp the sensor in one hand and, with the other hand, unscrew the
cable connector counter-clockwise at the sensor.
2) Pull out the cable connector plug from the expired sensor.
3) Unscrew the flow diverter from the sensor and discard the expired
sensor.
Note: The sensor contains lead and lead acetate, be sure to dispose
of expired sensors in accordance with hospital, local, state and
federal regulations.
4) Remove the new sensor from the packaging and remove the
protective film from the sensor face.
5) Insert the cable connector plug into the receptacle of the new sensor
and tighten the cable connector.
6) Screw the flow diverter onto the new sensor.
7) Wait approximately 20 minutes for the sensor to reach equilibrium.
8) Calibrate the new sensor.
Note: If the monitor is on when the sensor is detached and replaced,
the monitor will automatically force a re-calibration. The display will
read “CAL”.