To avoid explosion, do not operate the Oxygen monitor in the presence of
flammable anesthetics or in an atmosphere of explosive gases. Operating
the oxygen monitor in flammable or explosive atmospheres may result in
fire or explosion.
Never allow an excess length of cable near the patient’s head or neck, as
such could result in strangulation. Secure excess cable to the bed rail or
suitable object.
Never use a MAXO
2 monitor with a cable that appears worn, cracked or
has damaged insulation
Before use, all individuals who will be using the MAXO2 must become
thoroughly familiar with the information contained in this Operation
Manual. Strict adherance to the operating instructions is necessary for
safe effective product performance. This product will perform only as
designed if installed and operated in accordance with the manufacturer’s
operating instructions.
Use only genuine Maxtec accessories and replacement parts. Failure to
do so may seriously impair the monitor’s performance. Repair or
alteration of the MAXO
2 beyond the scope of the maintenance
instructions or by anyone other than an authorized Maxtec service person
could cause the product to fail to perform as designed.
Calibrate the MAXO
2 weekly when in operation and if enviromental
conditions change significantly. (ie, Temperature, Humidity, Barometric
Pressure. --- Refer to Calibration section of this manual).
Use of the MAXO
2 near devices that generate electrical fields may cause
erratic readings.
If the MAXO
2 is ever exposed to liquids (from spills or immersion) or to
any other physical abuse, turn the instrument OFF and then ON. This will
allow the unit to go through its self test and make sure everything is
operating correctly.
Never autoclave, immerse or expose the MAXO
2 (including sensor) to
high temperatures (>70
o
C). Never expose the device to pressure,
irradiation vacuum, steam, or chemicals.
To protect the unit from potential leaky battery damage always remove
batteries when the unit is going to be stored (not in use for 1 month) and
replace dead batteries with recognized name brand AA Alkaline batteries.