Data Sheet

SPECIFICATION
MODEL
DC VOLTAGE
CURRENT
Peak(10 sec.)
Convection
Convection
RATED
POWER
OUTPUT
LINE REGULATION
VOLTAGE ADJ. RANGE
LOAD REGULATION
SETUP, RISE TIME
HOLD UP TIME
(Typ.)
VOLTAGE RANGE
Note.4
FREQUENCY RANGE
EFFICIENCY
(Typ.)
INPUT
INRUSH CURRENT
(Typ.)
ISOLATION RESISTANCE
WORKING HUMIDITY
WORKING TEMP.
OVER TEMPERATURE
STORAGE TEMP.
TEMP. COEFFICIENT
VIBRATION
MTBF
DIMENSION (L*W*H)OTHERS
NOTE
PACKING
OVERLOAD
OVER VOLTAGE
AC CURRENT
(Typ.)
600ms, 30ms/230VAC 600ms, 30ms/115VAC at full load
15ms/230VAC 15ms/115VAC at full load
80 ~ 264VAC 113 ~ 370VDC
47 ~ 63Hz
92%91% 93%
16.5 ~ 19.5V
94%
26.4 ~ 31.2V
93.5%
29.7 ~ 35V 52.8 ~ 62.4V
2.3A/115VAC 1.1A/230VAC
COLD START 30A/115VAC 60A/230VAC
130~160% rated output power
13.2 ~ 15.6V
Protection type : Hiccup mode, recovers automatically after fault condition is removed
20 ~ 90% RH non-condensing
-30 ~ +85 (Refer to "Derating Curve")℃
-40 ~ +85℃
±℃ ℃0.03%/ (0 ~ 50 )
10 ~ 500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes
468Khrs min. MIL-HDBK-217F (25 )℃
76.2*50.8*28mm
or 3" * 2" *1.1" inch
±1.0%
±0.5%
±1.0%
±0.5%
±1.0%
±0.5% ±0.5%
±1.0% ±1.0%
±0.5%
±2.0% ±2.0% ±1.0% ±1.0% ±1.0%
100mVp-p
11.4~12.6V 14.3~15.8V 22.8~25.2V 25.6 ~ 28.4V 45.6 ~50.4V
120mVp-p 150mVp-p 150mVp-p 200mVp-p
15V 24V 27V 48V
11.8A
9.5A
9.5A
7.6A
6.25A
5A
5.55A
4.44A
3.125A
2.5A
141.6W
114W 114W
142.5W
120W
150W
119.9W
149.8W
120W
150W
RPS-120S-12 RPS-120S-15 RPS-120S-24 RPS-120S-27 RPS-120S-48
SAFETY STANDARDS
RIPPLE & NOISE (max.)
Note.2
VOLTAGE TOLERANCE Note.3
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1 f & 47 f parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. Derating may be needed under low input voltages. Please check the derating curve for more details.
5. Touch current was measured from primary input to DC output.
6. The ambient temperature derating of 3.5 /1000m with fanless models and of 5 /1000m with fan models for operating altitude higher than 2000m(6500ft).℃℃
7. The power supply is considered a component which will be installed into a final equipment. All the EMC tests are been executed by
mounting the unit on a mm* mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets 360 360
EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies.
(as available on http://www.meanwell.com)
ENVIRONMENT
SAFETY &
EMC
(Note 7)
PROTECTION
Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP, Secondary-Earth:1xMOPP
IEC60601-1, TUV EN60601-1, , UL ANSI / AAMI ES60601-1 (3.1 version),EAC TP TC 004
CAN/CSA-C22.2 No. 60601-1:14 - Edition 3 approved;
Design refer to EN60335 1 (By request)-
12V
LEAKAGE CURRENT(max.) Note.5
Earth leakage current < 150 A/264VAC , touch current < 80 A/264VAC
WITHSTAND VOLTAGE
ISOLATION RESISTANCE
I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC
I/P-O/P, I/P-FG, O/P-FG: 100M Ohms / 500VDC / 25 / 70% RH℃
Protection type : Shut down o/p voltage, re-power on to recover
Protection type :
S o/phut down voltage, recovers automatically after temperature goes down
Parameter
Parameter
ESD
Standard
Standard
EN55011 (CISPR11)
EN55011 (CISPR11)
EN61000-3-2
EN61000-3-3
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
EN61000-4-11
Test Level / Note
Test Level / Note
Class B
Class A
-----
EMC IMMUNITY
EMC EMISSION
EN60601-1-2
95% dip 1 periods, 30% dip 25 periods,
95% interruptions 250 periods
OPERATING ALTITUDE
Note.6
4000 meters
μ
μ μ
RF field susceptibility
EFT bursts
Surge susceptibility
Conducted susceptibility
Magnetic field immunity
Voltage dip, interruption
Conducted emission
Radiated emission
Harmonic current
Voltage flicker
Level 4, 15KV air ; Level 4, 8KV contact
Level 4, 30A/m
Level 3, 10V
Level 3, 2KV
Level 4, 4KV/Line-FG; 2KV/Line-Line
Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
Peak(10 sec.)
μ
POWER FACTOR PF>0.94/230VAC PF>0.98/115VAC at full load
0.13Kg; 100pcs/14Kg/1.13CUFT
120W 3"×2" Reliable Green Medical Power Supply
RPS-120S series
Class : Class B ClassⅠⅡ, Class : A
File Name:RPS-120S-SPEC 2019-05-27