Datasheet

SPECIFICATION
MODEL
DCVOLTAGE
RATEDCURRENT
CURRENTRANGE
RATEDPOWER
OUTPUT
VOLTAGE ADJ.RANGE
SETUP,RISETIME
HOLDTIME (Typ.)
VOLTAGERANGE
FREQUENCY RANGE
EFFICIENCY(Typ.)
ACCURRENT (Typ.)
INPUT
INRUSHCURRENT (Typ.)
LEAKAGECURRENT
SAFETY STANDARDS
WORKINGTEMP.
WORKINGHUMIDITY
STORAGETEMP.,HUMIDITY
TEMP.COEFFICIENT
VIBRATION
MTBF
DIMENSIONOTHERS
NOTE
PACKING
OVERLOAD
500ms,30ms/230VAC500ms,30ms/115VACatfullload
100ms/230VAC20ms/115VACatfullload
90~264VAC127~370VDC
47~63Hz
1.5A/115VAC1A/230VAC
COLDSTART 25A/115VAC50A/230VAC
140~180% ratedoutputpower
-20~+70 (Refertooutputloadderatingcurve)
20~90%RH non-condensing
-40~+85 ,10~95%RH
0.03%/ (0~45 )
10~500Hz,2G10min./1cycle,periodfor 60min.eachalongX, Y,Zaxes
FileName:RPT-75-SPEC2006-12-18
OVERVOLTAGE
HARMONICCURRENT
EMSIMMUNITY
RPT- 7 5 se ri es
RIPPLE&NOISE(max.) Note.2
VOLTAGETOLERANCE Note.3
EMICONDUCTION&RADIATION
ENVIRONMENT
SAFETY &
EMC
(Note4)
PROTECTION
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives.
5. Length of set up time is measured at cold first start. Turning ON/OFF the power supply may lead to increase of the set up time.
Protectiontype: Hiccupmode,recoversautomaticallyafterfaultconditionisremoved
Protectiontype:Shutdowno/pvoltage,re-powerontorecover
LINEREGULATION
LOADREGULATION
CH1CH1OUTPUTNUMBER
Features :
Universal AC input/Full range
Low leakage current<200uA
Protections:Short circuit/Over load/Over voltage
Free air convection for rated power and 23.5CFM forced air
convection for peak load
UL60601-1/IEC60601-1/EN60601-1 medical safety approved
Fixed switching frequency at 65KHz
3 years warranty
75W TripleOutputMedical Type
127*76.2*31mm(L*W*H)
CompliancetoEN55011(CISPR11),EN55022(CISPR22)ClassB
Complianceto EN61000-3-2,-3
I/P-O/P,I/P-FG,O/P-FG:100MOhms/500VDC
UL60601-1,TUVEN60601-1 Approved
I/P-O/P:4KVACI/P-FG:1.5KVACO/P-FG:1.5KVAC
PEAKLOAD(23.5CFM)
WITHSTANDVOLTAGE
ISOLATIONRESISTANCE
521.2Khrsmin.MIL-HDBK-217F(25 )
0.26Kg;63pcs/17.4Kg/1.35CUFT
CH1:4.75~5.5V ------
RPT-7503RPT-75D
CH2CH2 CH3CH3
12V12V24V 3.3V5V
1A
0~1.5A0.1~1A
1.5A
0~8A0.1~2A
5A
0.7~7A0.6~7A
61.8W73W
81W95W
120mVp-p200mVp-p 120mVp-p120mVp-p 80mVp-p80mVp-p
6.0%8.0% +10,-6%8.0%
1.0%1.0% 1.5%1.0%
+5,-4%
3.0% 6.0%3.0%
4.0%2.0%
1.0%0.5%
+3,-4%
1.5%
74%79%
CH1:5.75~6.75V CH1:3.8~4.45V
5V
1A 6A6A
CompliancetoEN61000-4-2,3,4,5,6,8,11;ENV50204,EN55024,EN60601-1-2,EN61000-6-2,EN61204-3heavyindustrylevel,EN61204-3Medicallevel,criteria A
Earthleakagecurrent<200uA /264VAC,Patient <100uA/264VACleakagecurrent
EN60601-1
IEC60601-1