Scanner User Manual

Pressure Systems, Inc. Model 9116 User’s Manual
Page 55 www.PressureSystems.com
Command ‘c’— Sub-command Index 03: Clear Stream(s)
This sub-command is used to “undefine” any previously configured host stream in a module,
one at a time, or all together. The sub-command’s format is:
Command
c 03 st
‘c’ is the command letter
‘03’ is the sub-command index (‘ii’) for configuration
st’ is the stream identifier character (1, 2, or 3 or 0=all streams)
NOTE: all parameters are separated by a space character.
Response
A
‘A’ is the acknowledge letter
Autonomous
Packet
none generated
Description: This sub-command clears (un-defines) the particular specified host stream (st=1-
3), or un-defines “all configured” host streams with a single command (st=0).
Once cleared, a stream must be reconfigured before it can be started.
Example:
Stop all the streams configured previously. Then clear (un-define) only stream 3.
Finally, resume the remaining defined streams 1 and 2:
c 02 0
c 03 3
c 01 0
Read response:
“A”
“A”
“A”