READ TRANSDUCER TEMPERATURE (Command ‘t’)
Purpose: Returns the most recently acquired engineering-unit temperature data (in ºC) for
the specified channels. Each datum returned in the response will be in the
specified high-precision data format.
Command
tppppf
‘t’ is the command letter
pppp’ is the position field
f’ is the format field
Response
dddd.. dddd
dddd’ are the datum fields, each with leading space (except f =7
or 8).
Description: The 4-character hex position field (pppp) specifies a 16-bit binary bit-map, with
each set bit (1) specifying a particular channel number (16-1, left-to-right).
The 1-character format field (f) specifies the format of each data field (dddd) that
will be returned in the requested response. The first datum returned in the
response will be for the highest channel number specified. Each (non-
binary) datum will be preceded by a space character. Some formats may not be
applicable to the specific type of data being requested. Valid formats are shown
in the following table:
f converts each internal response datum value from . . . max. char.
0 single binary float to 7-10-digit signed decimal “ [-xxx]x.xxxxxx” 13
1 single binary float to 8-digit hex integer “ xxxxxxxx” 9
2 double binary float to 16-digit hex integer “ xxxxxxxxxxxxxxxx” 17
5 single binary float to long integer (EU*1000) then to 8-digit hex integer 9
7 single binary float to single binary float (big endian: msb first) 4
8 single binary float to single binary float (little endian: lsb first) 4