W W W. M E A S U R E T E K . N E T PS-102 User/Technical Manual Contents subject to change without notice Version 1.
TABLE OF CONTENTS 1. INTRODUCTION and INSTALLATION ..................................................................................................................... 1 General and Safety Information ............................................................................................................................................................... 1 Specifications.....................................................................................................................................
1. INTRODUCTION and INSTALLATION General and Safety Information Risk of Electrical Shock: Disconnect all power sources before making cable connections to the scale platform or indicator. For use in dry environments only. Do not operate in hazardous areas. Read and understand all operating instructions before using this product. Keep this manual for future reference. Record the weight shortly after placing a load on the platform.
Contents Indicator AC120V/DC6V 500mA UL adapter Quick Guide & Technical Manual Base 6’ USB Cable Connection types 4 pins socket used: Pin1: Excitation + Pin2: Signal + Pin3: Excitation – Pin4: Signal - (4 holes socket) Indicator (4 pins socket) Base Installation 1. 2. 3. 4. Take the base and indicator out of the package, connect the indicator cable plug to the base cable plug. Place the scale on a hard, flat, and level surface. Adjust the feet to level the scale.
Indicator Display >0<- Scale is zeroed and gross weight is 0, tare is 0. Net - Display reading is net weight; tare is not 0. lb, kg, oz - Unit of measure. Hold - Scale is in dynamic weighing mode. - Hold flashes - actual fluctuating weight is displayed. - Hold does not flash - locked weight is displayed.
4. 5. 6. 7. To change the weight unit of measure, press the UNIT key. To send data to another device via the serial port, press the PRINT/HOLD key. To hold the weight data, press and hold the PRINT/HOLD key for 4 seconds. Power off the scale by pressing and holding the ZERO/ON/OFF key for 4 seconds. ZERO If the display dose not shows 0, and there is nothing on the platter, press the ZERO/ON/OFF key to zero the reading. Zero range: ±20% * full Capacity.
displayed. The default standard weight is 100% FS. Load 10%-100%FS (this must be equal or larger than that for the second calibration point) weight on the scale. Use HOLD or UNIT key to input the standard weight’s value. Press the TARE key to confirm. The indicator will flash the input weight. If the indicator receives reasonable data, it will be directed to the next step automatically. If an error occurred, the scale will display “CAL.Er” and return back to step3 for re-calibration. 6. When “CAL.
6. Configuration Parameters Setup 1. When the scale is off, press and hold the ZERO/ON/OFF and PRINT/HOLD keys until ‘C0NF’ is shown, which indicates that the scale is in Configuration parameter setup mode. 2. During setup mode, press the UNIT key to change the flashed digits, and use the HOLD/PRINT key to shift the flashed position. Press the TARE key to confirm the flashed digits. Press the ZERO/ON/OFF key to exit this mode. 3.
C8 2 Parameter Default C9 0 C10 0 C11 6 C12 2 C13 5 C14 1 C15 0 C16 0 Zero range for ZERO/ON/OFF key: 0 –power-on zero -point ±1%FS; 1 - power-on zero –point ±2%FS; 2 - power-on zero –point ±3%FS; 3 - power-on zero –point ±4%FS; 4 - power-on zero –point ±5%FS; 5 - power-on zero –point ±10%FS; 6 - power-on zero –point ±20%FS; 7 - power-on zero –point ±50%FS; 8 - power-on zero –point ±100%FS; 9 - no limitation; Setting Select which zero point will be used after scale is powered on and curr
Table1: use Kg as calibration unit: Calibration division value 0.0001kg 0.001kg 0.01kg 0.1kg 1kg 10kg 0.0002kg 0.002kg 0.02kg 0.2kg 2kg 20kg 0.0005kg 0.005kg 0.05kg 0.5kg 5kg 50kg Table2: use LB as calibration unit: Calibration division value 0.0001lb 0.001 lb 0.01 lb 0.1 lb 1 lb 10 lb 0.0002 lb 0.002 lb PS‐102 Display division value in different weight unit that can be used kg lb lb:oz 0.0001kg 0.0002lb Not available 0.001kg 0.002lb Not available 0.01kg 0.02lb 0.5oz 0.1kg 0.
0.02 lb 0.2 lb 2 lb 20 lb 0.0005 lb 0.005 lb 0.05 lb 0.5 lb 5 lb 50 lb 0.01 kg 0.1 kg 1 kg 10 kg 0.0002 kg 0.002 kg 0.02 kg 0.2 kg 2 kg 20 kg 0.02 lb 0.2 lb 2 lb 20 lb 0.0005 lb 0.005 lb 0.05 lb 0.5 lb 5 lb 50 lb 0.5 oz 5 oz Not available Not available Not available 0.1 oz 1 oz Not available Not available Not available 4. More Information for Configuration Parameters: 4.1 The division (d) of scale is determined by C3,C4 and C5 : If C3=3 (interval is 5), C4=2(decimal point is x.
7. User Parameters Setup In normal weighing mode, press UNIT and ZERO/ON/OFF until “USer” is shown to enter into the user parameters setup mode. 2. During this mode, press the UNIT key to change the flashed digits, press the HOLD/PRINT key to shift the flashed position. Press the TARE key to confirm and save the set data and enter next setting. Press the ZERO/ON/OFF key to exit this mode. 3.
4. 5. 4=continuously output gross, tare, net weight, unit and status data, data has not been received; 5=output weight, unit and status data one time when scale becomes stable; 6=output gross, tare, net weight, unit and status data one time when scale becomes stable; 7=Command –response mode.
The following details contain more information when U5 is set to 7: a) b) c) d) The baud rate and data format is set by U6 and U7. Responses to serial commands will be immediate, or within one weight measure cycle of the scale. One second is adequate for use as a time-out value by remote (controlling) device. The length of the weight field will be 8 digit weight data, one for minus sign, one for decimal point, two for measure unit (e.g. “lb”, “kg”).
(4). (5). (6). (7). (8). e) f) Response: H1H2H3 Zero function is activated and then it returns to current scale status, similar to pressing ZERO key. If ZERO function cannot be activated, it will return to current scale status. Command: T (54h 0dh) Response: H1H2H3 TARE function is activated, and it returns to current scale status, similar to pressing the TARE key. If TARE function cannot be activated, it will return to current scale status.
4 5 6 7 always 1 always 1 always 0 parity always 1 always 1 always 1 Parity always 1 always 1 always 0 parity 8. Symbol Definitions: 0﹉﹉﹉ -Zero point is over the setting range 0﹍﹍﹍ -Zero is below the setting range Ad﹉﹉﹉ -Analog digital converter chip over max. range Ad﹍﹍﹍ -Analog digital converter chip below min. range ﹉﹉﹉﹉ -Weight signal is too large ﹍﹍﹍﹍ -Weight signal is too small EEP.E1 -Config parameters incorrect (no set, no calibration, over normal range, etc.) EEP.
0﹉﹉﹉ Indication is out of key zero range 0﹍﹍﹍ Weight reading below Power On Zero limit ﹉﹉﹉﹉ ﹍﹍﹍﹍ SYMPTOM EEP.E1 EEP.E2 CAL.Er Cannot zero the display Reduce the weight on platform until the indication is within the key zero range. 1. Install platform on the scale. 2. Loosen the shipping protection screws. 3. Check whether an object is stuck between scale base, if yes, remove the object. 4. Perform zero calibration 1. Weight reading exceeds Overload limit; 1.
11.One Year Limited Warranty MeasureTek products covered in this manual are guaranteed to be free from defects in material and workmanship for a period of one year after date of purchase. Misuse, accidental damage, overload, alteration, and improper installation are expressly excluded.