Use Instructions

49
18. International regulations
18.1 Electromagnetic compatibility (EMC)
Breast pumps are intended to be used in a hospital or home environment. The Freestyle Flex
breast pump is EMC-tested in conformity with the requirements of IEC 60601-1-2:2007 3rd
Edition and IEC 60601-1-2:2014 4th Edition according to clause 7 and 8.9.
WARNING
The electric breast pump Freestyle Flex should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the electric breast pump Freestyle Flex
should be observed to verify normal operation in the configuration in which it will be used.
Wireless communications equipment such as wireless home network devices, mobile phones,
cordless telephones and their base stations, walkie-talkies can affect the electric breast pump
and should be kept at a distance of at least 30 cm away from the device.
18.2 Radio transmission
FCC (USA) statement
Changes or modifications not expressly approved by the party responsible for compliance
could void the users authority to operate the equipment.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two
conditions:
(1) this device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that may cause
undesired operation.
ISED (Canada) statement
This device complies with Industry Canada’s licence-exempt RSSs. Operation is subject to the
following two conditions:
(1) This device may not cause interference; and
(2) This device must accept any interference, including interference that may cause undesired
operation of the device.
Le présent appareil est conforme aux CNR d’Industrie Canada applicables aux appareils
radio exempts de licence. Lexploitation est autorisée aux deux conditions suivantes :
1) l’appareil ne doit pas produire de brouillage;
2) l’appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est
susceptible d’en compromettre le fonctionnement.
18.3 Drinking equipment
Parts intended for infant feeding comply with the European Standard EN 14350.