User's Manual

Specifications
Medical measurement
Measurement method Oscillometric measurement
Measurement range Pressure: 20 – 280 mmHg
Pulse: 40 – 200 beats / minute
Accuracy
Pressure: +/- 3 mmHg or 2%, whichever is
greater
Pulse: +/- 5%
Radio Frequency communication
Frequency
433.92 MHz (EU)
866.3 MHz (EU)
915 MHz (USA)
Max transmission power
10 dbm (EIRP) at 433.92 MHz and 866.3
MHz.
-1 dbm (EIRP) at 915 MHz
Bandwidth
60 kHz max.
Transmission range
200 m (free space)
Power Supply
Batteries 4 x 1.5v AA batteries
Environment
Operating condition +10˚C to +40˚C / 30% RH to 85% RH
Storage condition -10˚C to +60˚C / 30% RH to 85% RH
Dimensions Approx. 147[W] x 64[H] x 110 [D] mm
Weight Approx. 300g, excluding batteries
Classification Type BF
Model BP 100