User's Manual

Rev. 05
1
Bluetooth feature has been activated.
Repair the device, making sure to use the proper device
ID.
Consistently low
temperatures detected
Ensure that the device has been properly placed (see
section 5, step 3).
Verify that the room temperature is within the acceptable
range (60.8°F to 104°F).
Warning! Low body temperature requires seeking
immediate medical attention.
Consistently high
temperatures detected
Ensure that the device has been properly placed (see
section 5, step 3).
Verify that the room temperature is within the acceptable
range (60.8°F to 104°F).
Warning! High fever requires seeking immediate medical
attention.
10) Specifications
1. Physical Characteristics
Model: WT-001
Watch Body Dimensions: 38.15 mm x 22.5 mm x 11.7 mm
Watch Body Weight: 7.6 g
Watch Strap Length: 85-160 mm (adjustable)
Case Material: ABS Plastic (inner shell), TPE (strap)
Probe Material: Polycarbonate (lens), Stainless Steel (spot probe)
2. Electrical Characteristics
Battery: Li Ion, 50 mAh, rechargeable
Battery Life: Approximately 1 week per charge
Charging Port Rating: 5 V, 2 A
Measurement Location: Wrist
Firmware Version: WT-SW-001
Measurement Frequency: Every 5 minutes
Measurement Range (Temperature): 95°F to 107.6°F (35°C to 42°C)
Accuracy (Temperature): Complies with ASTM E 1112-00 standard requirements
Temperature Scale: Celsius or Fahrenheit
3. Environmental Conditions
Operation Environment Temperature: 60.8°F to 104°F (16°C to 40°C)
Operation Environment Humidity: 15%-95% humidity (non-condensing)
Operation Environment Pressure: 4 psi to 16 psi (74 kPa to 110 kPa)
Storage & Transport Environment Temperature: 14°F to 113°F (-10°C to 45°C)
Storage & Transport Environment Humidity: 15%-95% humidity (non-condensing)
Storage Environment Pressure: 4 psi to 16 psi (74 kPa to 110 kPa)