User Manual

10 11
OPERATING FUNCTIONS
1. LOW PRESSURE ALARM:
a) This device has both a visual and audible low pressure alarm built in. The
device constantly monitors the internal pressures of the sleeves to avoid
a pressure malfunction from occurring.
b) If a low pressure is detected, the pump will send a visual and audio alarm
warning after two cycles.
c) Use troubleshooting guide to fix the low-pressure alarm.
2. INDICATOR LIGHT:
a) Green on Power button: Illuminated to indicate power on with AC power.
b) Amber: Flashing to indicate a fault has been detected follow by an audio
alarm.
CLEANING & DISINFECTION
CAUTION
• Do not use phenol based cleaning solutions.
• Switch off the electrical supply to the pump and disconnect the
power cord from the main supply before cleaning and inspection.
The pump: The casing of the pump is manufactured from ABS plastic. If soiled it can
be wiped down with a sodium hypochlorite solution to dilution of 1000ppm or any
EPA approved, hospital grade cleaner.
The Sleeve: It is a single patient use sleeve. Do not wash or clean, discard after use.
The control unit(face) should also be cleaned weekly using a damp soft cloth and
mild detergent. (Switch off the electrical supply to the pump and disconnect the
power cord from the main supply before cleaning and inspection).
WASTE DISPOSAL
This Product has been supplied from an environmentally aware manu-
facturer that complies with the WEEE.
This product may contain substances that could be harmful to the
environment if disposed of in places (landfills) that are not appropriate
according the legislation. Please be environmentally responsible and
recycle this product through your recycling facility at its end of life.
ERROR TROUBLESHOOTING GUIDE
Indicates the error information for Hemo-Force Lite, these include:
Error Code /
Fault Type
Description Troubleshooting
Low Pressure • Leg is less than 40mmHg
pressure; foot pressure is less
than 120mmHg after
2 consecutive cycles.
• Air hose disconnected
at sleeves.
• Check Sleeves application
to ensure proper fit.
• Check for leaks in the sleeves or air
hose set and replace if faulty.
• Turn the pump off and restart.
• Refer to a service professional if
problem persists.
Sleeve Error • No sleeves attached during
start up.
• Sleeves are not fully deflated
during each restart.
• The initialization failed due to
a possible blockage in air hose/
connectors.
• Attach sleeves to pump and restart.
• Ensure sleeves are fully deflated
before restart.
• Check connectors for any blockage
and restart.
• Check air hose connection.
• Refer to a service professional if
problem persists.
HI Pressure/
System Failure
• Air hose is kinked.
• Air hose is disconnected at air
outlet of the pump.
• Pressure is greater than set
pressure after two consecutive
inflations.
• Pressure is lower than set
pressure after 10 consecutive
inflations.
• Electrical malfunction.
• Check air hose set for any kinks.
• Attach sleeves to pump and restart.
• Turn the pump off and restart.
• Refer to a professional for service if
problem persists.