User Manual

16 17
Guidance and manufacturer’s declaration – electromagnetic immunity (cont)
The Hemo-Force Lite is intended for use in the electromagnetic environment (for home healthcare)
specified below. The customer or the user of the Hemo-Force Lite should assure that it is used in such an
environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic
environment – guidance (for home
healthcare environment)
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms:
in ISM and amateur
radio bands
between 0,15 MHz
and 80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms:
in ISM and amateur
radio bands
between 0,15 MHz
and 80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Portable and mobile RF communica-
tions equipment should be used no
closer to any part of the Hemo-Force
Lite, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1,2
d = 1,2 80 MHz to 800 MHz
d = 2,3 800 MHz to 2,7 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the recommend-
ed separation distance in metres (m).
Interference may occur in the
vicinity of equipment marked
with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Guidance and manufacturer’s declaration – electromagnetic immunity Test specifications
for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
The Hemo-Force Lite is intended for use in the electromagnetic environment (for home healthcare)
specified below. The customer or the user of the Hemo-Force Lite should assure that it is used in such an
environment.
Test
frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
power
(W)
Distance
(m)
IMMUNITY
TEST
LEVEL
(V/m)
Compliance
LEVEL
(V/m)
(for home
health-
care)
385
380-
390
TETRA 400
Pulse
modulation b)
18 Hz
1,8 0,3 27 27
450
430-
470
GMRS 460,
FRS 460
FM c) ±5 kHz
deviation
1 kHz sine
2 0,3 28 28
710
704-
787
LTE Band 13,
17
Pulse
modulation b)
217 Hz
0,2 0,3 9 9745
780
810
800-
960
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation b)
18 Hz
2 0,3 28 28
870
930
1720
1700-
1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band
1, 3,
4, 25; UMTS
Pulse
modulation b)
217 Hz
2 0,3 28 281845
1970
2450
2400-
2570
Bluetooth,
WLAN,
802.11
b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation b)
217 Hz
2 0,3 28 28
5240
5100-
5800
WLAN
802.11
a/n
Pulse
modulation b)
217 Hz
0,2 0,3 9 95500
5785
NOTE: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting
antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is
permitted by IEC 61000-4-3.