Product Manual

Vital Signs Monitor User’s Manual
13-3
Attention:
Excessive or frequent cleaning may damage airbag, so don’t clean airbag
unless necessary.
Do not dry airbag and sheath in high temperature.
If higher sterilization level is required, please choose disposal cuff.
One disposal cuff can only be used for one patient.
Carefully keep water and cleaning solution out of the connecting parts ofcuff
and monitor.
13.3.3 CO2 Sensor and Reusable Airway Adapter
The outside of the module or sensor may be cleaned and disinfected by wiping
with 70% isopropyl alcohol, a 10% bleach solution, or mild soap. After
cleaning, wipe with a clean, water-dampened cloth to rinse. Dry before use.
Reusable airway adapters may be cleaned by rinsing in a warm soapy solution,
followed by soaking in a liquid disinfectant such as 70% isopropyl alcohol, a
10% bleach solution, Cidex® or System (refer to the disinfectant
manufacturer's instructions for use). Adapters should then be rinsed with
sterile water and dried.
Reusable airway adapters may also be pasteurized or autoclaved. Autoclave at
121
℃ (250℉) for 20 minute
s, unwrapped.
Before reusing the adapter, ensure the windows are dry and residue-free, and
that the adapter has not been damaged during handling or by the cleaning
process.
13.3.4 Temp probe
The recommended disinfector is isopropyl alcohol 70%. Don’t use undiluted
decolourant (5%5.25% sodium hypochlorite) or other non-recommended
disinfector in order to avoid damage to sensor.
Please clean the probe lens as the following steps:
1. Use the pily stick or soft cloth bedewed with water or alcohol to wipe the lens gently. Do
not pot the probe in the water or other liquid.
2. Dry the lens for over 30 min before cover the new earcap, and avoid solarizing directly
while the temperature is beyond 10℃~40.