Part 1: PPC/Pump system initialization
35
Set alarms
Alarms alert the user in the event the PPC or Pump recognizes an insulin
delivery problem. The Alarm Feedback screen must always be in the
“ON” position.
1. Press ACT to enter the “ALARMS”
menu.
2. The PPC has three alarm options, two
audible tones (Low/High) and a vibrate
mode. Press the
and
buttons to
select the desired alarm, then press ACT.
3. The screen will now display “SET
ALARM FEEDBACK”. This setting
should always be “ON”. Press ACT.
4. Place the PPC near the Pump. When the
communication is completed, the PPC
screen will change to “SELF TEST” and
then to the Time/Date screen.
NOTE: When the alarm type is set
to “vibrate” the beeps from the PPC during nor-
mal programming will be low volume.
ALARMS
SET
PPC
ALARM TYPE
LOW/HIGH/VIBRATE
SET
ALARM
FEEDBACK
ON/OFF
PPC
COMMUNICATING