Installation Instructions Part 1
Table Of Contents
- Table of Contents
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System 11
- CHAPTER 2 Indications and Contraindications 21
- CHAPTER 3 Personal Pump Communicator (PPC) 25
- CHAPTER 4 Pump Implantation 77
- CHAPTER 5 Pump Refill Procedure 97
- CHAPTER 6 Explanting the Pump System 107
- CHAPTER 7 Warnings And Precautions 109
- CHAPTER 8 Adverse Reactions 115
- CHAPTER 9 System Alarms and Messages 117
- CHAPTER 10 Troubleshooting Pump System Under-delivery 131
- CHAPTER 11 MiniMed 2007C Implantable Insulin Pump System 133
- APPENDIX A Label Information 137
- APPENDIX B Implant Worksheet 141
- APPENDIX C Refill Form 143
- APPENDIX D Precautions and General Procedures 145
- APPENDIX E Pump Rinse Procedure 151
- APPENDIX F Side Port Catheter Flush Procedure 163
- APPENDIX G Stroke Volume Measurement 175
- APPENDIX H Pressure Measurement Using The Side Port Catheter 181
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System
- CHAPTER 2 Indications and Contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- Introduction
- Part 1: PPC/Pump system initialization
- Part 2: Additional PPC programming features
- Main menu
- Program a bolus
- Set a normal bolus with the variable bolus feature turned off
- Set a normal bolus with the variable bolus feature turned on
- Set a square wave bolus
- Set a dual wave bolus
- Review bolus history
- Suspend mode
- Programming a basal rate
- Programming basal delivery pattern
- Setting basal rate profiles in each delivery pattern
- Program a temporary basal rate
- Set a temporary basal rate
- Stop a temporary basal rate
- Personal events
- History
- Main menu
- Setup Pump
- Setup II
- Audio bolus
- Activating the audio bolus feature
- Set an audio bolus from the main menu bolus screen
- Variable bolus
- Refill
- Priming
- Diagnostic rate
- Initialize to factory defaults
- Download software
- Stop Pump
- Supervisor password
- Exit supervisor
- Personal Pump communicator messages
- Clinical history codes
- Clinical history PPC/Pump codes
- Audio bolus
- CHAPTER 4 Pump Implantation
Personal Pump Communicator (PPC)
36
Set maximum bolus, basal rate and time display format
This programming is performed in the “SETUP II” menus. These screens
allow healthcare professionals to limit the maximum amount of insulin a
patient can deliver, either when taking a bolus or setting a new basal rate.
Access to the “SETUP II” menus is through the “SETUP PUMP” screen.
1. Press SEL until the “SETUP PUMP”
screen is displayed. Press ACT and press
SEL to reach the “SETUP II” screen.
Press ACT and then SEL to reach the
“MAX BOLUS” screen.
2. Press ACT and the maximum bolus
amount (units) will start flashing. Press
the ▲
▲▲
▲ and ▼
▼▼
▼ buttons to change the maxi-
mum allowable bolus (0.0 to 25.0 units)
and then press ACT again.
3. Place the PPC near the Pump and com-
plete the communication process. The
PPC screen will automatically change to
the “MAX BASAL” screen.
4. Press ACT and the screen will change to
“SET MAX BASAL RATE”. The maxi-
mum basal amount will start flashing.
Press the ▲
▲▲
▲ and ▼
▼▼
▼ buttons to change the
maximum allowable basal rate (0.2 to
35.0 units/hour) and then press ACT
again.
5. Place the PPC near the Pump and com-
plete the communication process. The
PPC screen will automatically change to
“TIME FORMAT.”
MAX BOLUS
--
u
MAX BOLUS
--- -----
u
SET
PPC
COMMUNICATING
MAX BASAL RATE
0.2u/h
SET
PPC
COMMUNICATING