Set maximum bolus, basal rate and time display format
This programming is performed in the “SETUP II” menus. These screens
allow healthcare professionals to limit the maximum amount of insulin a
patient can deliver, either when taking a bolus or setting a new basal rate.
Access to the “SETUP II” menus is through the “SETUP PUMP” screen.
1. Press SEL until the “SETUP PUMP”
screen is displayed. Press ACT and press
SEL to reach the “SETUP II” screen.
Press ACT and then SEL to reach the
“MAX BOLUS” screen.
2. Press ACT and the maximum bolus
amount (units) will start flashing. Press
the
and
buttons to change the maxi-
mum allowable bolus (0.0 to 25.0 units)
and then press ACT again.
3. Place the PPC near the Pump and com-
plete the communication process. The
PPC screen will automatically change to
the “MAX BASAL” screen.
4. Press ACT and the screen will change to
“SET MAX BASAL RATE”. The maxi-
mum basal amount will start flashing.
Press the
and
buttons to change the
maximum allowable basal rate (0.2 to
35.0 units/hour) and then press ACT
again.
5. Place the PPC near the Pump and com-
plete the communication process. The
PPC screen will automatically change to
“TIME FORMAT.”
MAX BOLUS
--
u
MAX BOLUS
--- -----
u
SET
PPC
COMMUNICATING
MAX BASAL RATE
0.2u/h
SET
PPC
COMMUNICATING