Installation Instructions Part 1
Table Of Contents
- Table of Contents
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System 11
- CHAPTER 2 Indications and Contraindications 21
- CHAPTER 3 Personal Pump Communicator (PPC) 25
- CHAPTER 4 Pump Implantation 77
- CHAPTER 5 Pump Refill Procedure 97
- CHAPTER 6 Explanting the Pump System 107
- CHAPTER 7 Warnings And Precautions 109
- CHAPTER 8 Adverse Reactions 115
- CHAPTER 9 System Alarms and Messages 117
- CHAPTER 10 Troubleshooting Pump System Under-delivery 131
- CHAPTER 11 MiniMed 2007C Implantable Insulin Pump System 133
- APPENDIX A Label Information 137
- APPENDIX B Implant Worksheet 141
- APPENDIX C Refill Form 143
- APPENDIX D Precautions and General Procedures 145
- APPENDIX E Pump Rinse Procedure 151
- APPENDIX F Side Port Catheter Flush Procedure 163
- APPENDIX G Stroke Volume Measurement 175
- APPENDIX H Pressure Measurement Using The Side Port Catheter 181
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System
- CHAPTER 2 Indications and Contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- Introduction
- Part 1: PPC/Pump system initialization
- Part 2: Additional PPC programming features
- Main menu
- Program a bolus
- Set a normal bolus with the variable bolus feature turned off
- Set a normal bolus with the variable bolus feature turned on
- Set a square wave bolus
- Set a dual wave bolus
- Review bolus history
- Suspend mode
- Programming a basal rate
- Programming basal delivery pattern
- Setting basal rate profiles in each delivery pattern
- Program a temporary basal rate
- Set a temporary basal rate
- Stop a temporary basal rate
- Personal events
- History
- Main menu
- Setup Pump
- Setup II
- Audio bolus
- Activating the audio bolus feature
- Set an audio bolus from the main menu bolus screen
- Variable bolus
- Refill
- Priming
- Diagnostic rate
- Initialize to factory defaults
- Download software
- Stop Pump
- Supervisor password
- Exit supervisor
- Personal Pump communicator messages
- Clinical history codes
- Clinical history PPC/Pump codes
- Audio bolus
- CHAPTER 4 Pump Implantation
Personal Pump Communicator (PPC)
40
Program a basal rate
1. From the Time/Date screen, press SEL
until the “BASAL RATE” screen is dis-
played. Preset delivery pattern “A”, a
basal rate of 0.2 U/H, and the word
“NOW” is flashing. Press ACT.
2. A “1” now appears to the right of the “A”
indicating that this programming will
effect the 1
st
basal change within the “A”
pattern, (there are 3 patterns available
[A,B,C] and 48 basal changes possible
within each pattern). The flashing
0.2 U/H indicates the value can be changed. Use the ▲
▲▲
▲ and ▼
▼▼
▼ but-
tons to change the value and then press ACT.
NOTE: 00:00 indicates a start time of MIDNIGHT in
24hr. display mode. In 12hr. display mode, the
screen indicates the start time as 12:00am.
3. The screen now displays “SET TIME”,
and a time of 00:30 or 12:30 am (24 or 12
hour respectively) and a “2.”
If one basal is all that will be used, press
ACT two times. If more than one basal
rate is to be programmed, enter a start
time and amount of the new basal rate for that time period, then press
ACT and enter the new basal rate.
The user can enter a new basal rate at 30 minute intervals, up to 48
basal rates.
4. Place the PPC near the Pump and com-
plete the communication process.
08:26 OCT 12
BASAL RATE : A NOW
00:00 0.2u/h
-- U/H
00:30
SET RATE
BASAL RATE: A
PPC
COMMUNICATING
Dmp9196021-011_c.book Page 40 Thursday, April 4, 2002 8:15 AM