Installation Instructions Part 1

Table Of Contents
Personal Pump Communicator (PPC)
52
5. A “3” now appears on the screen with
“SET TIME.” Follow the same procedure
previously described and program a new
profile. If no additional profiles are
needed change the flashing time to dashes
(by pressing
) and press ACT.
6. The screen will indicate “PPC COMMU-
NICATING.” Place the PPC near the
Pump and complete the communication
process.
7. The screen will briefly display the calcu-
lated 24 hour basal dose based on the
basal rate programming. In this example
a total of “8.4U” will be delivered.
To set multiple basal profiles in the other pat-
terns (A,B,C), select the pattern in SETUP II menu and follow the same
procedure.
Program a temporary basal rate
A Temporary Basal Rate is often used when a brief change in basal deliv-
ery is required, for example during exercise.
Set a temporary basal rate
1. From the time and date display press SEL
until the “TEMPORARY BASAL”
screen is displayed.
2. Press ACT and the “SET DURATION”
screen appears. The time duration of the
Temporary Basal Rate is displayed as
flashing dashes. Press the
and
but-
tons to enter a time duration
, in 30 minute
increments from 30 minutes to 24 hours.
BASAL RATE : 3
SET TIME
05:00 0.2u/h
PPC
COMMUNICATING
24 HOUR TOTAL
8.4u
TEMP BASAL
--- --- --- --- u/h
TEMP BASAL
--- --- u/h
00:30
SET DURATION