Personal Pump Communicator (PPC)
70
7. Place the PPC over the Pump and com-
plete the communication process.
8. The PPC now displays the “PUMP
RESET” screen, and beeps six times
every minute until the Pump is restarted.
Press SEL then ACT. The PPC returns
to the Time/Date screen. The Pump may
now be reprogrammed.
Download software
This feature has been disabled at the factory.
1. From the “PUMP REFILL” screen in the
Supervisor menu, press SEL until the
“DOWNLOAD SOFTWARE” screen
appears.
Stop Pump
This screen allows the Physician to stop the operation of the Pump.
From the “PUMP REFILL” screen in the
Supervisor Menu, press SEL until the “STOP
PUMP” screen appears.
Press ACT and the word “CONFIRM”
appears highlighted with the entire screen
flashing. Press ACT again to confirm.
PPC
COMMUNICATING
08:13 OCT 02
PUMP
RESET
DOWNLOAD
SOFTWARE
STOP PUMP
STOP PUMP
CONFIRM