Setup II
71
Place the PPC near the Pump and allow the
communication to complete.
The PPC will beep four times to confirm the
“STOP PUMP” command. To restart the
Pump, press SEL and “PUMP STOPPED”
will start to flash. Press ACT. Place the PPC
near the Pump. When the communication is
complete, the PPC returns to the Time/Date
screen.
Supervisor password
If the password is lost, access the Supervisor Menu as follows :
Program the time to midnight (12:00 am or 00:00) and the date to
January 01.
Enter the password 0000 within one minute. If the password is
not entered within one minute the time will be reset to midnight
again.
Record the new password in the patient’s chart.
PPC
COMMUNICATING
08:13 OCT 02
PUMP STOPPED