Pump Implantation
78
Supplies and solutions
Before preparing the Pump, be sure to read Appendix D, Precautions and
General Procedures. Assemble the necessary materials prior to starting
the procedure:
Supplies
One (1) 100 µL sterile pipette (available non sterile from
MiniMed) MMT-4104
One (1) sterile scalpel blade
One (1) scale (0.01g resolution)
One (1) sterile beaker (or sterile barrier towel) for weighing
One (1) sterile bag for PPC
One (1) Pump System: PPC, Pump and Side Port Catheter
Sterile markers and Steri-Strips
®
Three (3) refill Kits - MMT-4105:
1 kit to remove shipping fluid from Pump
1 kit to fill the Pump with U-400 INSULIN (to rinse the
pump)
1 kit to fill the Pump with U-400 INSULIN (final insulin
fill)
One (1) 3 mL syringe (fluid barrier)
Three (3) MiniMed refill needles, MMT-4102
Three (3) 18 gauge needles
One (1) Implant Worksheet
One (1) Back-up Pump System: PPC, Pump and Side Port Cathe-
ter
Solutions
One (1) bottle of sterile water, room temperature
One (1) vial 10 mL rinse buffer (RB)
Four (4) vials 10 mL Aventis HOE 21 PH U-400 INSULIN