Installation Instructions Part 1
Table Of Contents
- Table of Contents
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System 11
- CHAPTER 2 Indications and Contraindications 21
- CHAPTER 3 Personal Pump Communicator (PPC) 25
- CHAPTER 4 Pump Implantation 77
- CHAPTER 5 Pump Refill Procedure 97
- CHAPTER 6 Explanting the Pump System 107
- CHAPTER 7 Warnings And Precautions 109
- CHAPTER 8 Adverse Reactions 115
- CHAPTER 9 System Alarms and Messages 117
- CHAPTER 10 Troubleshooting Pump System Under-delivery 131
- CHAPTER 11 MiniMed 2007C Implantable Insulin Pump System 133
- APPENDIX A Label Information 137
- APPENDIX B Implant Worksheet 141
- APPENDIX C Refill Form 143
- APPENDIX D Precautions and General Procedures 145
- APPENDIX E Pump Rinse Procedure 151
- APPENDIX F Side Port Catheter Flush Procedure 163
- APPENDIX G Stroke Volume Measurement 175
- APPENDIX H Pressure Measurement Using The Side Port Catheter 181
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System
- CHAPTER 2 Indications and Contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- Introduction
- Part 1: PPC/Pump system initialization
- Part 2: Additional PPC programming features
- Main menu
- Program a bolus
- Set a normal bolus with the variable bolus feature turned off
- Set a normal bolus with the variable bolus feature turned on
- Set a square wave bolus
- Set a dual wave bolus
- Review bolus history
- Suspend mode
- Programming a basal rate
- Programming basal delivery pattern
- Setting basal rate profiles in each delivery pattern
- Program a temporary basal rate
- Set a temporary basal rate
- Stop a temporary basal rate
- Personal events
- History
- Main menu
- Setup Pump
- Setup II
- Audio bolus
- Activating the audio bolus feature
- Set an audio bolus from the main menu bolus screen
- Variable bolus
- Refill
- Priming
- Diagnostic rate
- Initialize to factory defaults
- Download software
- Stop Pump
- Supervisor password
- Exit supervisor
- Personal Pump communicator messages
- Clinical history codes
- Clinical history PPC/Pump codes
- Audio bolus
- CHAPTER 4 Pump Implantation
Personal Pump Communicator (PPC)
38
Lock maximum bolus/basal, enter personal ID and password, stop
Pump
To access the Supervisor Mode press SEL until the “SETUP PUMP”
screen is displayed. Then press and hold down the ▲
▲▲
▲ and ▼
▼▼
▼ buttons
simultaneously until the “ENTER SUPERVISOR PASSWORD” screen
appears.
Patients should not be given the Supervisor
Mode password, to avoid the accidental
programming of a large priming bolus
(99.8 U) or diagnostic insulin rate.
1. The first zero will be flashing. Press the
▲
▲▲
▲ and ▼
▼▼
▼ buttons to select the first digit,
then press ACT.The screen advances to
the second zero. Press the ▲
▲▲
▲ and ▼
▼▼
▼ but-
tons to select the second digit, then press
ACT. Repeat for the last two digits. The
factory pre-set password is YIQ8.
2. Entry into the Supervisor Mode is indi-
cated by the screen “PUMP REFILL.”
3. Press SEL until the “SET MAXIMUMS
SCREEN” is displayed, and then ACT to
reach “SET MAXIMUMS”.
Press the ▲
▲▲
▲ and ▼
▼▼
▼ buttons to select
“ON” if the patient is not
given access to
this feature, or “OFF” if the patient is
per-
mitted access. Press ACT again.
!
WARNING
ENTER
SUPERVISOR
PASSWORD
0000
PUMP
REFILL
SET
MAXIMUMS OFF/ON
Dmp9196021-011_c.book Page 38 Thursday, April 4, 2002 8:15 AM