Installation Instructions Part 1
Table Of Contents
- Table of Contents
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System 11
- CHAPTER 2 Indications and Contraindications 21
- CHAPTER 3 Personal Pump Communicator (PPC) 25
- CHAPTER 4 Pump Implantation 77
- CHAPTER 5 Pump Refill Procedure 97
- CHAPTER 6 Explanting the Pump System 107
- CHAPTER 7 Warnings And Precautions 109
- CHAPTER 8 Adverse Reactions 115
- CHAPTER 9 System Alarms and Messages 117
- CHAPTER 10 Troubleshooting Pump System Under-delivery 131
- CHAPTER 11 MiniMed 2007C Implantable Insulin Pump System 133
- APPENDIX A Label Information 137
- APPENDIX B Implant Worksheet 141
- APPENDIX C Refill Form 143
- APPENDIX D Precautions and General Procedures 145
- APPENDIX E Pump Rinse Procedure 151
- APPENDIX F Side Port Catheter Flush Procedure 163
- APPENDIX G Stroke Volume Measurement 175
- APPENDIX H Pressure Measurement Using The Side Port Catheter 181
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System
- CHAPTER 2 Indications and Contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- Introduction
- Part 1: PPC/Pump system initialization
- Part 2: Additional PPC programming features
- Main menu
- Program a bolus
- Set a normal bolus with the variable bolus feature turned off
- Set a normal bolus with the variable bolus feature turned on
- Set a square wave bolus
- Set a dual wave bolus
- Review bolus history
- Suspend mode
- Programming a basal rate
- Programming basal delivery pattern
- Setting basal rate profiles in each delivery pattern
- Program a temporary basal rate
- Set a temporary basal rate
- Stop a temporary basal rate
- Personal events
- History
- Main menu
- Setup Pump
- Setup II
- Audio bolus
- Activating the audio bolus feature
- Set an audio bolus from the main menu bolus screen
- Variable bolus
- Refill
- Priming
- Diagnostic rate
- Initialize to factory defaults
- Download software
- Stop Pump
- Supervisor password
- Exit supervisor
- Personal Pump communicator messages
- Clinical history codes
- Clinical history PPC/Pump codes
- Audio bolus
- CHAPTER 4 Pump Implantation
Part 1: PPC/Pump system initialization
39
4. Place the PPC near the Pump and com-
plete the communication process. The
PPC screen will automatically advance to
the “PERSONAL ID” screen. Press
ACT.
5. The first of the 32 possible ID locations is
flashing. Enter the patient ID (alpha-
numeric) by pressing the ▲
▲▲
▲ and ▼
▼▼
▼ but-
tons and then ACT after each entry. Con-
tinue to press ACT, activating each “0”
until the screen changes.
6. Place the PPC near the Pump and com-
plete the communication process.
7. Press SEL until the “SUPERVISOR
PASSWORD” screen is displayed. Then
press ACT.
8. The screen now reads, “SET SUPERVI-
SOR PASSWORD”. Press ACT. Use the
▲
▲▲
▲ and ▼
▼▼
▼ buttons to enter a new supervi-
sor password (alphanumeric), pressing
ACT after each entry.
NOTE: Record the password in the patient’s chart.
PPC
COMMUNICATING
SET PERSONAL ID
..................................
................
.20KOLO5
PPC
COMMUNICATING
SET
SUPERVISOR
PASSWORD
Y1Q8
SET
SUPERVISOR
PASSWORD
0000
Dmp9196021-011_c.book Page 39 Thursday, April 4, 2002 8:15 AM