Installation Instructions Part 1

Table Of Contents
Personal Pump Communicator (PPC)
44
Set a normal bolus with the variable bolus feature turned on
1. From the Time/Date screen press SEL
until the “BOLUS” screen is displayed.
The last bolus value programmed and the
Time and Date will be flashing.
2. Press ACT and the “SET BOLUS TYPE”
screen appears. If “NORMAL” is not
flashing, use the
and
buttons to
select “NORMAL.” Press ACT.
3. The “SET BOLUS” screen appears, with
dashes under “IMM” flashing. Use the
and
buttons to enter an immediate
bolus amount.
4. Press ACT and the “CONFIRM” screen
is displayed, with the screen flashing.
Confirm the bolus amount by pressing
ACT again.
5. Place the PPC near the Pump and com-
plete the communication process.
6. When the bolus programming is com-
plete, the PPC will beep once and then
briefly display the amount of insulin cur-
rently delivered.
The Pump will beep at each of the first five strokes (if audio feedback is
ON). The PPC beeps at the end of the bolus. Three segments of the insulin
delivery icon will be displayed and spinning slowly during the bolus
delivery. By pressing SEL you can read the amount of insulin delivered.
08:13 Jan 02
BOLUS
IMM EXT
-- U -- U
SET
NORMAL
BOLUS TYPE
SET BOLUS
IMM
--- --- U
CONFIRM
IMM
2.6u
PPC
COMMUNICATING
08:23 JAN 04
0.0U
BOLUS
Dmp9196021-011_c.book Page 44 Thursday, April 4, 2002 8:15 AM