Installation Instructions Part 1
Table Of Contents
- Table of Contents
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System 11
- CHAPTER 2 Indications and Contraindications 21
- CHAPTER 3 Personal Pump Communicator (PPC) 25
- CHAPTER 4 Pump Implantation 77
- CHAPTER 5 Pump Refill Procedure 97
- CHAPTER 6 Explanting the Pump System 107
- CHAPTER 7 Warnings And Precautions 109
- CHAPTER 8 Adverse Reactions 115
- CHAPTER 9 System Alarms and Messages 117
- CHAPTER 10 Troubleshooting Pump System Under-delivery 131
- CHAPTER 11 MiniMed 2007C Implantable Insulin Pump System 133
- APPENDIX A Label Information 137
- APPENDIX B Implant Worksheet 141
- APPENDIX C Refill Form 143
- APPENDIX D Precautions and General Procedures 145
- APPENDIX E Pump Rinse Procedure 151
- APPENDIX F Side Port Catheter Flush Procedure 163
- APPENDIX G Stroke Volume Measurement 175
- APPENDIX H Pressure Measurement Using The Side Port Catheter 181
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System
- CHAPTER 2 Indications and Contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- Introduction
- Part 1: PPC/Pump system initialization
- Part 2: Additional PPC programming features
- Main menu
- Program a bolus
- Set a normal bolus with the variable bolus feature turned off
- Set a normal bolus with the variable bolus feature turned on
- Set a square wave bolus
- Set a dual wave bolus
- Review bolus history
- Suspend mode
- Programming a basal rate
- Programming basal delivery pattern
- Setting basal rate profiles in each delivery pattern
- Program a temporary basal rate
- Set a temporary basal rate
- Stop a temporary basal rate
- Personal events
- History
- Main menu
- Setup Pump
- Setup II
- Audio bolus
- Activating the audio bolus feature
- Set an audio bolus from the main menu bolus screen
- Variable bolus
- Refill
- Priming
- Diagnostic rate
- Initialize to factory defaults
- Download software
- Stop Pump
- Supervisor password
- Exit supervisor
- Personal Pump communicator messages
- Clinical history codes
- Clinical history PPC/Pump codes
- Audio bolus
- CHAPTER 4 Pump Implantation
Part 2: Additional PPC programming features
51
Setting basal rate profiles in each delivery pattern
Each of the basal delivery patterns is a set of up to 48 basal rates, one for
each half-hour of the day.
1. Press SEL until the “BASAL RATE”
screen is displayed. A basal pattern is
selected (for example Pattern A). Press
ACT.
2. A “SET RATE” and “1” is now displayed
to the right of the “A” indicating this pro-
gramming will set the 1
st
basal rate
within the “A” profile. The “0.2U/H” is
now flashing, indicating the value can be
changed. Use the ▲
▲▲
▲ and ▼
▼▼
▼ arrow keys
to program a new value, for example “0.4 U/H”, and then press ACT.
NOTE: 00:00 indicates a start time of MIDNIGHT in 24hr.
display mode. 12:00am indicates a start time of
MIDNIGHT in 12hr. display mode.
3. The screen now displays “SET TIME”,
with a time of “00:30” or “12:30am”
flashing (24 or 12 hour respectively) and
a “2.” This screen allows the second basal
rate to be set. Enter a start time for the 2
nd
basal rate within the “A” profile, for
example “04:30.” Press ACT. (Example: a second basal rate of 0.4U/
H starting at 04:30.)
4. This screen changes to “SET RATE”
again, indicating the 2
nd
basal rate can
now be programmed. Use the ▲
▲▲
▲ and ▼
▼▼
▼
arrow buttons to enter a new rate, for
example “0.2U/H”, and then press ACT.
BASAL RATE : A NOW
0.2U/H
00:00
BASAL RATE : A
0.2U/H
00:00
SET RATE
BASAL RATE : A 2
SET TIME
04:30 0.4u/h
BASAL RATE : 2
SET TIME
04:30 0.4u/h