Installation Instructions Part 1
Table Of Contents
- Table of Contents
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System 11
- CHAPTER 2 Indications and Contraindications 21
- CHAPTER 3 Personal Pump Communicator (PPC) 25
- CHAPTER 4 Pump Implantation 77
- CHAPTER 5 Pump Refill Procedure 97
- CHAPTER 6 Explanting the Pump System 107
- CHAPTER 7 Warnings And Precautions 109
- CHAPTER 8 Adverse Reactions 115
- CHAPTER 9 System Alarms and Messages 117
- CHAPTER 10 Troubleshooting Pump System Under-delivery 131
- CHAPTER 11 MiniMed 2007C Implantable Insulin Pump System 133
- APPENDIX A Label Information 137
- APPENDIX B Implant Worksheet 141
- APPENDIX C Refill Form 143
- APPENDIX D Precautions and General Procedures 145
- APPENDIX E Pump Rinse Procedure 151
- APPENDIX F Side Port Catheter Flush Procedure 163
- APPENDIX G Stroke Volume Measurement 175
- APPENDIX H Pressure Measurement Using The Side Port Catheter 181
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System
- CHAPTER 2 Indications and Contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- Introduction
- Part 1: PPC/Pump system initialization
- Part 2: Additional PPC programming features
- Main menu
- Program a bolus
- Set a normal bolus with the variable bolus feature turned off
- Set a normal bolus with the variable bolus feature turned on
- Set a square wave bolus
- Set a dual wave bolus
- Review bolus history
- Suspend mode
- Programming a basal rate
- Programming basal delivery pattern
- Setting basal rate profiles in each delivery pattern
- Program a temporary basal rate
- Set a temporary basal rate
- Stop a temporary basal rate
- Personal events
- History
- Main menu
- Setup Pump
- Setup II
- Audio bolus
- Activating the audio bolus feature
- Set an audio bolus from the main menu bolus screen
- Variable bolus
- Refill
- Priming
- Diagnostic rate
- Initialize to factory defaults
- Download software
- Stop Pump
- Supervisor password
- Exit supervisor
- Personal Pump communicator messages
- Clinical history codes
- Clinical history PPC/Pump codes
- Audio bolus
- CHAPTER 4 Pump Implantation
Personal Pump Communicator (PPC)
52
5. A “3” now appears on the screen with
“SET TIME.” Follow the same procedure
previously described and program a new
profile. If no additional profiles are
needed change the flashing time to dashes
(by pressing ▼
▼▼
▼
) and press ACT.
6. The screen will indicate “PPC COMMU-
NICATING.” Place the PPC near the
Pump and complete the communication
process.
7. The screen will briefly display the calcu-
lated 24 hour basal dose based on the
basal rate programming. In this example
a total of “8.4U” will be delivered.
To set multiple basal profiles in the other pat-
terns (A,B,C), select the pattern in SETUP II menu and follow the same
procedure.
Program a temporary basal rate
A Temporary Basal Rate is often used when a brief change in basal deliv-
ery is required, for example during exercise.
Set a temporary basal rate
1. From the time and date display press SEL
until the “TEMPORARY BASAL”
screen is displayed.
2. Press ACT and the “SET DURATION”
screen appears. The time duration of the
Temporary Basal Rate is displayed as
flashing dashes. Press the ▲
▲▲
▲ and ▼
▼▼
▼ but-
tons to enter a time duration
, in 30 minute
increments from 30 minutes to 24 hours.
BASAL RATE : 3
SET TIME
05:00 0.2u/h
PPC
COMMUNICATING
24 HOUR TOTAL
8.4u
TEMP BASAL
--- --- --- --- u/h
TEMP BASAL
--- --- u/h
00:30
SET DURATION
Dmp9196021-011_c.book Page 52 Thursday, April 4, 2002 8:15 AM