Installation Instructions Part 1
Table Of Contents
- Table of Contents
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System 11
- CHAPTER 2 Indications and Contraindications 21
- CHAPTER 3 Personal Pump Communicator (PPC) 25
- CHAPTER 4 Pump Implantation 77
- CHAPTER 5 Pump Refill Procedure 97
- CHAPTER 6 Explanting the Pump System 107
- CHAPTER 7 Warnings And Precautions 109
- CHAPTER 8 Adverse Reactions 115
- CHAPTER 9 System Alarms and Messages 117
- CHAPTER 10 Troubleshooting Pump System Under-delivery 131
- CHAPTER 11 MiniMed 2007C Implantable Insulin Pump System 133
- APPENDIX A Label Information 137
- APPENDIX B Implant Worksheet 141
- APPENDIX C Refill Form 143
- APPENDIX D Precautions and General Procedures 145
- APPENDIX E Pump Rinse Procedure 151
- APPENDIX F Side Port Catheter Flush Procedure 163
- APPENDIX G Stroke Volume Measurement 175
- APPENDIX H Pressure Measurement Using The Side Port Catheter 181
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System
- CHAPTER 2 Indications and Contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- Introduction
- Part 1: PPC/Pump system initialization
- Part 2: Additional PPC programming features
- Main menu
- Program a bolus
- Set a normal bolus with the variable bolus feature turned off
- Set a normal bolus with the variable bolus feature turned on
- Set a square wave bolus
- Set a dual wave bolus
- Review bolus history
- Suspend mode
- Programming a basal rate
- Programming basal delivery pattern
- Setting basal rate profiles in each delivery pattern
- Program a temporary basal rate
- Set a temporary basal rate
- Stop a temporary basal rate
- Personal events
- History
- Main menu
- Setup Pump
- Setup II
- Audio bolus
- Activating the audio bolus feature
- Set an audio bolus from the main menu bolus screen
- Variable bolus
- Refill
- Priming
- Diagnostic rate
- Initialize to factory defaults
- Download software
- Stop Pump
- Supervisor password
- Exit supervisor
- Personal Pump communicator messages
- Clinical history codes
- Clinical history PPC/Pump codes
- Audio bolus
- CHAPTER 4 Pump Implantation
Personal Pump Communicator (PPC)
58
Setup Pump
The “SETUP PUMP” screen permits access to the primary Setup menus
for the Pump. Setup features discussed previously in this chapter are ref-
erenced here. Other Setup features not previously discussed are presented
here. Press SEL until the “SETUP PUMP” screen is displayed, then press
ACT to access the SETUP menus.
Auto off
Auto Off is a safety feature, reminding the user to update insulin medica-
tion delivery programming in the Pump. An alarm can be set after a
period of programming inactivity, from one to 16 hours. At the onset of
the alarm the Pump will automatically be placed in SUSPEND mode.
1. From the “ SETUP PUMP” screen, press
ACT. Press SEL to reach the “AUTO
OFF” screen, then press ACT. The screen
will display “AUTO OFF” with flashing
dashes/time. Use the ▲
▲▲
▲ and ▼
▼▼
▼ buttons to
select the number of hours before an
“AUTO OFF” alarm occurs.
2. Press ACT. In this example, a time dura-
tion of 10 hours was selected. The PPC
will alarm if the user did not attempt to
program the Pump during the previous 10
hours and be placed in SUSPEND mode.
3. Place the PPC near the Pump and com-
plete the communication process.The
PPC will beep once and return to the
Time/Date screen.
AUTO OFF
--- --- HRS
SET
AUTO OFF
10 HRS
PPC
COMMUNICATING
Dmp9196021-011_c.book Page 58 Thursday, April 4, 2002 8:15 AM