Refill
The refill procedure will be described in Chapter 5 “Pump Refill Proce-
dure.” The following screens are used at the end of this procedure, and
allow you to enter the new refill volume and to calculate the refill accu-
racy.
The "PUMP REFILL" screen appears when
you access the Supervisor Mode. Press ACT.
The PPC prompts for the residual amount of
insulin removed from the Pump. Use the
and
buttons to enter the weight (grams) of
extracted insulin.
Press ACT and the PPC prompts for the refill
insulin amount. Use the
and
buttons to
enter the weight (grams) of insulin refilled in
the Pump.
Press ACT. Confirm that the insulin concen-
tration defaults to U-400. Press ACT again.
Place the PPC near the Pump to complete the
process.
The calculated accuracy will be displayed
briefly. Record this value on the refill form.
PUMP REFILL
SET
EXTRACTED AMOUNT
---- ---- g
SET
REFILL AMOUNT
---- ---- g
SET
INSULIN U400
CONCENTRATION
PPC
COMMUNICATING
CALCULATED
ACCURACY
---- ---- %