Installation Instructions Part 1

Table Of Contents
Setup II
67
Diagnostic rate
A diagnostic rate is intended for use only when
the Pump is filled with buffer.
The Diagnostic Rate feature is used to help “diagnose” Pump delivery
problems. Fill the Pump with rinse buffer before using this feature. To
access this feature, the Pump must be in the “PUMP STOPPED” mode.
1. From the “PUMP STOPPED” screen,
press SEL. The screen will begin flash-
ing.
2. Press the
and
buttons at the same
time until the “ENTER SUPERVISOR
PASSWORD” screen appears. Re-enter
the supervisor password, pressing ACT
after each letter or number is entered.
The factory preset password is: YIQ8.
3. When the Supervisor Password is suc-
cessfully entered, the screen will change
to “PUMP REFILL”. Press SEL until the
“DIAGNOSTIC RATE” screen appears.
4. Press ACT and the word “SET” appears
flashing. Use the
and
buttons to
enter a Diagnostic Rate (units/hour).
5. Press ACT and the word “SET” changes
to “DELIVER”. “NO” also appears
flashing. Use the
and
buttons
change “NO” to “YES”, then press ACT.
!
WARNING
08:13 OCT 12
PUMP STOPPED
0000
SET
SUPERVISOR
PASSWORD
--- --- u/h
DIAGNOSTIC
RATE
10 u/h
SET
DIAGNOSTIC
RATE
10 u/h
DELIVER
DIAGNOSTIC
RATE