Installation Instructions Part 1
Table Of Contents
- Table of Contents
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System 11
- CHAPTER 2 Indications and Contraindications 21
- CHAPTER 3 Personal Pump Communicator (PPC) 25
- CHAPTER 4 Pump Implantation 77
- CHAPTER 5 Pump Refill Procedure 97
- CHAPTER 6 Explanting the Pump System 107
- CHAPTER 7 Warnings And Precautions 109
- CHAPTER 8 Adverse Reactions 115
- CHAPTER 9 System Alarms and Messages 117
- CHAPTER 10 Troubleshooting Pump System Under-delivery 131
- CHAPTER 11 MiniMed 2007C Implantable Insulin Pump System 133
- APPENDIX A Label Information 137
- APPENDIX B Implant Worksheet 141
- APPENDIX C Refill Form 143
- APPENDIX D Precautions and General Procedures 145
- APPENDIX E Pump Rinse Procedure 151
- APPENDIX F Side Port Catheter Flush Procedure 163
- APPENDIX G Stroke Volume Measurement 175
- APPENDIX H Pressure Measurement Using The Side Port Catheter 181
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System
- CHAPTER 2 Indications and Contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- Introduction
- Part 1: PPC/Pump system initialization
- Part 2: Additional PPC programming features
- Main menu
- Program a bolus
- Set a normal bolus with the variable bolus feature turned off
- Set a normal bolus with the variable bolus feature turned on
- Set a square wave bolus
- Set a dual wave bolus
- Review bolus history
- Suspend mode
- Programming a basal rate
- Programming basal delivery pattern
- Setting basal rate profiles in each delivery pattern
- Program a temporary basal rate
- Set a temporary basal rate
- Stop a temporary basal rate
- Personal events
- History
- Main menu
- Setup Pump
- Setup II
- Audio bolus
- Activating the audio bolus feature
- Set an audio bolus from the main menu bolus screen
- Variable bolus
- Refill
- Priming
- Diagnostic rate
- Initialize to factory defaults
- Download software
- Stop Pump
- Supervisor password
- Exit supervisor
- Personal Pump communicator messages
- Clinical history codes
- Clinical history PPC/Pump codes
- Audio bolus
- CHAPTER 4 Pump Implantation
Setup II
67
Diagnostic rate
A diagnostic rate is intended for use only when
the Pump is filled with buffer.
The Diagnostic Rate feature is used to help “diagnose” Pump delivery
problems. Fill the Pump with rinse buffer before using this feature. To
access this feature, the Pump must be in the “PUMP STOPPED” mode.
1. From the “PUMP STOPPED” screen,
press SEL. The screen will begin flash-
ing.
2. Press the ▲
▲▲
▲ and ▼
▼▼
▼ buttons at the same
time until the “ENTER SUPERVISOR
PASSWORD” screen appears. Re-enter
the supervisor password, pressing ACT
after each letter or number is entered.
The factory preset password is: YIQ8.
3. When the Supervisor Password is suc-
cessfully entered, the screen will change
to “PUMP REFILL”. Press SEL until the
“DIAGNOSTIC RATE” screen appears.
4. Press ACT and the word “SET” appears
flashing. Use the ▲
▲▲
▲ and ▼
▼▼
▼ buttons to
enter a Diagnostic Rate (units/hour).
5. Press ACT and the word “SET” changes
to “DELIVER”. “NO” also appears
flashing. Use the ▲
▲▲
▲ and ▼
▼▼
▼ buttons
change “NO” to “YES”, then press ACT.
!
WARNING
08:13 OCT 12
PUMP STOPPED
0000
SET
SUPERVISOR
PASSWORD
--- --- u/h
DIAGNOSTIC
RATE
10 u/h
SET
DIAGNOSTIC
RATE
10 u/h
DELIVER
DIAGNOSTIC
RATE
Dmp9196021-011_c.book Page 67 Thursday, April 4, 2002 8:15 AM