Installation Instructions Part 1

Table Of Contents
Pump Implantation
80
Remove shipping fluid from the Pump
1. Using aseptic technique, open the Pump sterile package. Do not
remove the plastic tubing from the Pump outlet. Trim the distal part
of the tubing with the scalpel blade.
2. Remove the refill syringe from the refill kit package.
3. Firmly attach the two-way stopcock to the refill syringe and attach an
18 gauge needle to the stopcock.
4. Draw approximately 4 mL of RINSE BUFFER into the refill syringe.
5. Fill the hub of the refill needle with RINSE BUFFER and firmly
attach it to the stopcock, prime the needle and close the stopcock.
Figure 6: Filling the Hub of the Refill Needle
Dmp9196021-011_c.book Page 80 Thursday, April 4, 2002 8:15 AM