Installation Instructions Part 1
Table Of Contents
- Table of Contents
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System 11
- CHAPTER 2 Indications and Contraindications 21
- CHAPTER 3 Personal Pump Communicator (PPC) 25
- CHAPTER 4 Pump Implantation 77
- CHAPTER 5 Pump Refill Procedure 97
- CHAPTER 6 Explanting the Pump System 107
- CHAPTER 7 Warnings And Precautions 109
- CHAPTER 8 Adverse Reactions 115
- CHAPTER 9 System Alarms and Messages 117
- CHAPTER 10 Troubleshooting Pump System Under-delivery 131
- CHAPTER 11 MiniMed 2007C Implantable Insulin Pump System 133
- APPENDIX A Label Information 137
- APPENDIX B Implant Worksheet 141
- APPENDIX C Refill Form 143
- APPENDIX D Precautions and General Procedures 145
- APPENDIX E Pump Rinse Procedure 151
- APPENDIX F Side Port Catheter Flush Procedure 163
- APPENDIX G Stroke Volume Measurement 175
- APPENDIX H Pressure Measurement Using The Side Port Catheter 181
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System
- CHAPTER 2 Indications and Contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- Introduction
- Part 1: PPC/Pump system initialization
- Part 2: Additional PPC programming features
- Main menu
- Program a bolus
- Set a normal bolus with the variable bolus feature turned off
- Set a normal bolus with the variable bolus feature turned on
- Set a square wave bolus
- Set a dual wave bolus
- Review bolus history
- Suspend mode
- Programming a basal rate
- Programming basal delivery pattern
- Setting basal rate profiles in each delivery pattern
- Program a temporary basal rate
- Set a temporary basal rate
- Stop a temporary basal rate
- Personal events
- History
- Main menu
- Setup Pump
- Setup II
- Audio bolus
- Activating the audio bolus feature
- Set an audio bolus from the main menu bolus screen
- Variable bolus
- Refill
- Priming
- Diagnostic rate
- Initialize to factory defaults
- Download software
- Stop Pump
- Supervisor password
- Exit supervisor
- Personal Pump communicator messages
- Clinical history codes
- Clinical history PPC/Pump codes
- Audio bolus
- CHAPTER 4 Pump Implantation
Pump Implantation
82
Rinse the Pump with insulin (IN1)
11. Remove the second refill syringe from the refill kit package.
12. Firmly attach the two-way stopcock to the refill syringe and attach an
18 gauge needle to the stopcock. Use the sterile marker to label the
syringe, “IN1”.
13. Draw 20 mL of INSULIN (two vials) into the refill syringe.
14. Remove the 18 gauge needle and expel all air bubbles in the syringe.
15. Close the stopcock.
16. Retract the plunger until it locks into place. This should be no further
than 55 mL. Do not go beyond the vent hole. Press the lock into the
plunger groove to be sure it is firmly secured.
17. Shake vigorously for a minimum of 30 seconds to degas the
INSULIN.
18. Point the syringe tip upward and slowly open the stopcock to vent the
syringe.
19. Release the locking ring on the refill syringe.
20. Expel air in the syringe and carefully observe to ensure no air bubbles
remain in the syringe. If air bubbles are noted, repeat steps 15 - 20.
21. Fill the hub of the refill needle with INSULIN and attach it to the
stopcock.
22. Prime the needle and close the stopcock.
23. Refill the fill port with the 3 mL RINSE BUFFER syringe, as needed,
to maintain the fluid barrier.
24. Use the “IN1” syringe containing the degassed INSULIN to fill the
Pump.
Dmp9196021-011_c.book Page 82 Thursday, April 4, 2002 8:15 AM