Preprogramming and pre-testing the pump
83
OPTIONAL
25. Use this step only if the plunger is not moving forward while filling
the reservoir.
26. With the refill needle pointing down, vent the syringe head space by
pulling back firmly on the plunger until the second sealing ring on the
black rubber cap passes beyond the vent hole (see Figure 7).
Figure 7: Venting the Syringe Head Space