CHAPTER 7 Warnings And Precautions
Warnings
The Physician should be completely familiar with the function of the
Pump, Catheter, and PPC prior to use. Patients should be provided a com-
plete copy of the Patient Manual and have demonstrated the ability to pro-
gram the PPC, recognize and respond to safety alarms, and take care of
the device prior to discharge.
The Implantable Insulin Pump can only be used with MMT-4027A and
4024A Side Port Catheters.
Only Aventis HOE 21 PH U-400 insulin may be used in the Medtronic
MiniMed 2007C Implantable Insulin Pump System. Use of other insulin
types may cause damage to the Pump mechanism resulting in impaired
insulin delivery or Pump failure.
Any unauthorized changes or modifications made to any component of
the Medtronic MiniMed 2007C Implantable Insulin Pump System may
prevent effective use of that and other components.
Electrotherapy
The Medtronic MiniMed 2007C Implantable Insulin Pump System has
been tested in close proximity with electrosurgical, electrocoagulation,
and cardiac defibrillation medical equipment. Typical use of this type of
device has not affected the Pump. However, patients should be instructed