Table 2: Preparation and Labeling of Syringes
Record patients blood glucose
Record the patient’s blood glucose value at the start of the procedure.
Monitor blood glucose every 30 minutes, or as needed.
Prepare syringes for emptying the Pump
Syringe 1 is used for emptying the Pump. Label and prepare syringe 1 per
Table 2. Special attention should be observed regarding the Precautions
and General Procedures section in Appendix D of this manual.
After the syringe is prepared, weigh the syringe and enter the weight on
the Refill Form.
Prepare syringes for filling the Pump
During the Catheter rinse/flush procedure, syringes 2, 3, 5, and 6 are used
for filling the Pump. Each syringe should be labeled with its solution and
syringe number, then filled and degassed per Table 2 and the General Pro-
cedures section in Appendix D of this manual.
Syringe # Syringe Type Volume Solution
#1 Refill 5 mL RINSE
BUFFER
#2 Refill 20 mL RINSE
BUFFER
#3 Refill 30 mL RINSE
BUFFER
#4 10ml or 20 ml 5-10 mL RINSE
BUFFER
#5 Refill 20 mL INSULIN
#6 Refill 30 mL INSULIN