Measuring Stroke Volume
1. In a sterile field immediately before the procedure; prepare the appa-
ratus. Prime the stopcock, pipette, and needle.
Figure 19: Measuring the Stroke Volume
2. Turn the stopcock to close off the needle.
3. Insert the needle into the side port.
4. Turn the stopcock. as shown in Figure 19.