Record patients blood glucose
Record the patient’s blood glucose value at the start of the procedure.
Monitor blood glucose every 30 minutes, or as needed.
Prepare syringes for emptying the pump
Syringe 1 is used for emptying the Pump. Label and prepare syringe 1 per
the Precautions and General Procedures section in Appendix D of this
manual. After the syringe is prepared, weigh it and record the weight on
the Refill Form.
Prepare syringes for filling the pump
During the pressure test procedure, syringes 2, 3, 5, and 6 are used for fill-
ing the Pump. Each of these syringes need to be labeled with its solution
and syringe number, and then filled and degassed per the General Proce-
dures in Appendix D of this manual.
Prepare syringe for priming the test setup
During the pressure test procedure, syringe 4 is used to prime the needle,
filter, and the 3-way stopcock. This syringe needs to be labeled and then
prepared by following the steps below:
1. Attach an 18 gauge needle to the syringe. Draw 5 - 10 ml of RINSE
BUFFER into the syringe.
2. Expel all air from the syringe.
3. Attach a 3-way stopcock to the syringe.
4. Turn the stopcock to close off the syringe.