Product Info Part 4

Table Of Contents
191
ing ACT on the “SUSPEND” screen.
Remove rinse buffer and fill with insulin
The RINSE BUFFER is removed from the Pump fluid system and the
system is filled with INSULIN. This procedure uses the following vol-
umes in syringes 3, 5, and 6:
Syringe 3 with residual RINSE BUFFER from the previous steps
Syringe 5 with 20 ml INSULIN
Syringe 6 with 30 ml INSULIN
Syringe 3: Rinse Buffer (residual)
1. Expel all air from the syringe and prime the needle.
2. Close the stopcock. Obtain a vacuum by pulling back on the plunger
until it locks. Press the lock into the plunger groove to be sure it is
firmly secured.
3. Enter the Pump with the refill needle and syringe.
4. Open the stopcock and remove the RINSE BUFFER. After the fluid
level stops moving, wait an additional 30 seconds to make sure all of
the RINSE BUFFER and air is removed.
5. When the time has elapsed, close the stopcock, remove the syringe,
and discard.
Syringe 5: Insulin
6. Prime the needle and close the stopcock.
7. Enter the Pump fill port with the refill needle.
8. Open the stopcock and allow the pump to fill completely with INSU-
LIN. When the fluid level stops moving, the Pump is filled.
9. Close the stopcock and remove the syringe.
10. Prepare syringe 6 for aspiration. Obtain a vacuum by pulling back on
Dmp9196021-011_c.book Page 191 Wednesday, April 3, 2002 5:07 PM