Product Info Part 4
Table Of Contents
- APPENDIX D Precautions and General Procedures
- Special note and precautions
- General procedures
- Locating the pump fill port and the side port
- Venting the Minimed refill syringe (optional)
- APPENDIX E Pump Rinse Procedure
- Preparing for the procedure
- Remove insulin from the Pump and fill with NaOH
- Equilibrate and pull NaOH through system
- APPENDIX F Side Port Catheter Flush Procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke Volume Measurement
- Supplies and Solutions
- APPENDIX H Pressure Measurement Using The Side Port Catheter
- Supplies and solutions
- Measuring pressure in the side port
153
Preparing for the procedure
NOTE: Before beginning any refill, flush, stroke volume
measurement, carefully read Appendix D, Precau-
tions and General Procedures, and keep this
in mind as you perform each procedure.
In order to prepare for the Side Port Catheter rinse and flush procedure, a
total of seven syringes will need to be labeled and then prepared with dif-
ferent solutions. Table 1 defines the syringe numbers and corresponding
solutions:
Prepare syringes for emptying the Pump
Syringe 1 is used for emptying the Pump. Label and prepare syringe 1 per
Table 1.
Prepare syringes for filling the Pump
During the Catheter rinse/flush procedure, syringes 2, 3, 4, 5, 6, and 7 are
used for filling the Pump. Each of these syringes needs to be labeled with
its solution and syringe number, and then filled and degassed per Table 1.
Program minimal basal rate
Place the PPC in a sterile bag and program the Pump to “SUSPEND”
mode, the basal rate will be 0.2 U/h.
Syringe # Syringe Type Volume Solution
Refill 5 ml RINSE BUFFER
Refill 20 ml NaOH
Refill 30 ml NaOH
Refill 20 ml RINSE BUFFER
Refill 30 ml RINSE BUFFER
Refill 20 ml INSULIN
Refill 30 ml INSULIN
Dmp9196021-011_c.book Page 153 Wednesday, April 3, 2002 5:07 PM