Programming Your Pump
2
7
Review the bolus history
You can review the bolus type, bolus amount, time and day of your last
512 insulin boluses.
1. From the Time/Date screen press SEL
until the BOLUS screen is displayed.
The last bolus value programmed and the
time and date will be flashing.
2. Press the
button once to display the
previous bolus and the time and date it was delivered. Each additional
press will display the next previous bolus delivery.
Suspend mode
Suspend Mode allows you to cancel bolus programming. In suspend
mode the pump will still deliver a basal rate of approximately 0.2u/hr.
1. From the Time/Date screen press SEL
until the SUSPEND PUMP screen is
shown.
2. Press ACT.
3. Place the PPC near the Pump to complete
the programming. When communication
is finished the Pump will beep 3 times.
4. All four segments of the insulin delivery
icon will be displayed.
5. To restart the Pump, press SEL and ACT.
08:23
Jan 04
BOLUS
--
IMM EXT
5.6
U
--
U
PROG
SUSPEND PUMP
PPC
COMMUNICATING
08:13
OCT 02
PUMP SUSPENDED