Programming Your Pump
44
Set a maximum basal rate
Your doctor will decide if you should use this feature. Setting a maximum
basal rate will protect you from over-delivering insulin in case you make
a programming mistake.
5. After setting a maximum bolus, the MAX BASAL screen appears
with the maximum basal rate flashing.
6. Use the
and
buttons to enter a maxi-
mum basal rate (0.2U/h to 35.0U/h), then
press ACT.
7. Place the PPC near the pump and com-
plete the communication. The PPC screen
will automatically change to the TIME
FORMAT screen.
Set time format
1. From the SETUP II screen, press SEL
until the SET TIME FORMAT screen
is displayed.
2. Use the
and
buttons to select either
a 12 hour (12:00am) or 24 hour (00.00)
time format, then press ACT.
3. Place the PPC near the pump and com-
plete the communication. The PPC screen
will change to the PERSONAL
EVENTS screen.
MAX BASAL RATE
0.2
U
/H
SET
PPC
COMMUNICATING
SET
TIME FORMAT
12/24 HOUR
PPC
COMMUNICATING