User Manual

Table Of Contents
Alarms and Messages
49
Pump alarms
Alarm feedback
The Alarm Feedback function allows you to verify the Pump and the
Pump beeper is operating normally. When Alarm Feedback is pro-
grammed YES, thePumpwillbeeponeachofthefirstfivePump
strokes:
Following a change in the delivery regimen, for example when
completing a meal Bolus and then changing to a Basal Rate.
After the Alarm Feedback function is programmed YES. Alarm
Feedback will stay on until programmed back to NO.
Pump low battery
The Implantable Insulin Pump battery is designed to last approximately
seven years during conditions of normal use (see Chapter 9, Technical
Specifications). Battery life may vary somewhat depending upon your
insulin delivery requirements. When battery energy becomes low, a volt-
age sensor in the Pump will trigger the Pump Low Battery Alarm. A
Pump Low Battery Alarm indicates there is approximately eight weeks of
battery energy remaining.
You can clear this condition and your Pump will continue to operate nor-
mally. However, you should notify your doctor as soon as possible. Your
doctor will then schedule a Pump replacement, and may switch you back
to conventional insulin therapy.
PPC Display
Type of
Alarm Pump Alarm
PUMP LOW BATTERY 1 In 24 hours if no PPC
communication