User Manual

Table Of Contents
Alarms and Messages
53
Low reservoir
When the Pump calculates that less than 800 units (2 ml) of insulin
remains in its reservoir, the following display will appear:
You can clear this message by pressing SEL and ACT, and then continue
programming. You should notify your doctor of the Low Reservoir mes-
sage, and schedule a Pump refill appointment as soon as possible.
Empty reservoir
When the Pump calculates that less that 400 units (1 ml) of insulin
remains in its reservoir, the following display will appear:
You can clear this message by pressing SEL and ACT, and then continue
programming. The PPC Medication Remaining function reads calculated
values only, and there may still be some insulin left in the Pump reservoir.
You should be report this alarm to your doctor as soon as possible, and
schedule an appointment for a Pump refill. It is important not to allow the
pump to deplete its insulin supply as this may result in a catheter block-
age.
PPC Display Type of Alarm Pump Alarm
LOW
RESERVOIR
1 In 24 hours if no PPC communication
PPC Display Type of Alarm Pump Alarm
EMPTY
RESERVOIR
1 In 24 hours if no PPC
communication